Chapter 5 Troubleshooting
EVIS EXERA TJF TYPE 160VR OPERATION MANUAL
Endoscope functions
Angulation
Air/water feeding
Irregularity
description
Possible cause
Solution
Resistance is
encountered when
rotating angulation
control knob(s).
The angulation lock(s) is
(are) engaged.
Rotate angulation lock(s) in the
“F
” direction.
Irregularity
description
Possible cause
Solution
No air feeding.
The air pump is not
operating.
Press the LOW, MED or HIGH
button on the light source as
described in the light source’s
instruction manual.
The air/water valve is
damaged.
Replace it with a new one.
No water feeding.
The air pump is not
operating.
Press the LOW, MED or HIGH
button on the light source as
described in the light source’s
instruction manual.
There is no sterile water
in the water container.
Add sterile water to fill the
container to the specified level.
The air/water valve is
damaged.
Replace it with a new one.
The air/water valve is
sticky.
The air/water valve is
dirty.
Remove the air/water valve.
Reprocess the air/water valve and
then attach it again.
The air/water valve is
damaged.
Replace it with a new one.
The air/water valve
cannot be attached.
An incorrect air/water
valve is used.
Use a correct air/water valve.
The air/water valve is
damaged.
Replace it with a new one.