38
Chapter 3 Preparation and Inspection
EVIS EXERA TJF TYPE 160VR OPERATION MANUAL
1.
If any ancillary equipment is ON, turn it OFF.
2.
Insert the endoscope connector completely into the scope socket (output
socket when using the CLV-U20/U40) of the light source.
3.
Connect the water container’s connection adapter to the air supply
connector and water supply connector (see Figure 3.25).
4.
Confirm that the water container’s connection adapter fits properly and that
the water container’s connection adapter cannot be rotated.
Figure 3.25
5.
Align the mark on the videoscope cable EXERA or the videoscope cable
100 with mark 1 on the endoscope connector and push it in until it stops
(see Figure 3.26).
Figure 3.26
Air supply connector
Water supply connector
Water container’s connection adapter
Endoscope connector
(1)
(2)
(3)
(4)
Mark 2 (yellow)
Mark (yellow)
Mark 1 (yellow)