Problem

Cause

Remedy

Page

e) Picture is too dark.

AE lock, which was set when the expo-
sure time was shorter than the currently
required exposure time, is active.

Cancel AE lock.

(Online
manual)

The output highlight level adjustment is
too low.

Reset the current level adjustment
and adjust the optimum level again.

(Online
manual)

The microscope illumination is too dark.

Increase the microscope illumination
intensity or disengage the existing ND
filter to increase brightness.

––

f ) The colors in the picture are

strange.

The area selected in white balance
adjustment was improper.

Select a white area as the rectangu-
lar white balance adjustment area.

(Online
manual)

The RGB balance is adjusted improp-
erly in manual white balance adjust-
ment.

Perform manual white balance adjust-
ment to adjust the RGB color balance
to obtain optimum colors.

(Online
manual)

The area selected in black balance
adjustment was improper.

Select a black area as the rectangu-
lar black balance adjustment area.

(Online
manual)

The screen color setting of the com-
puter is incorrect.

Set the computer display color to 24-
bit color or higher. The recommended
setting is 32-bit color.

––

g) The picture is not in focus.

The microscope is not focused prop-
erly

Adjust the focus correctly with the fine
adjustment knob.

––

The aperture iris diaphragm of the con-
denser is open too wide.

Close the aperture iris diaphragm a
little.

––

The field iris diaphragm is not set prop-
erly.

Adjust the field iris diaphragm until the
image circumscribes the field of view.

––

Lens components of the microscope are
contaminated or the cover glass on the
front of the camera is stained.

Clean the objective, photography lens,
condenser and/or window lens of the
microscope, or clean the cover glass
on the bottom of the camera head.

6

The microscope and/or camera are sub-
jected to vibration during recording.

Record images in an environment in
which the microscope and camera
are not vibrated. It is effective to use
an anti-vibration bench.

––

h) The 4140 x 3096 and 2070 x 1548

images are not neat.

The camera is subjected to vibration
during recording.

Record images in an environment in
which the microscope and camera
are not vibrated. It is effective to use
an anti-vibration bench.

––

i ) The DP2-TWAIN or DP2-BSW win-

dow is not displayed correctly or
the menu characters are not dis-
played correctly.

The resolution setting of the screen is
incorrect.

Set the resolution setting at 1280 x
1024 or more in the property of the
screen.

––

The large font has been selected for the
font size of the screen.

Select a small font in the property of
the screen.

––

j ) External trigger signal is not out-

put.

External triggering is not enabled.

Enable external triggering on the DP2-
BSW.

(Online
manual)

k) Still image cannot be recorded

using the external trigger input.

External triggering is not enabled.
} DP2-BSW Ver. 2.1 is not compatible

with the trigger input.

Enable external triggering on the DP2-
BSW.

(Online
manual)

Содержание DP72

Страница 1: ...familiarize yourself fully with the use of this camera we recommend that you study this manual thoroughly before operating the camera For image operations including recording editing and saving please...

Страница 2: ...the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a...

Страница 3: ...THE POWER SUPPLY CORD 33 34 1 6 7 8 9 10 11 15 CONTENTS IMPORTANT Be sure to read this chapter for safe use of the equipment DP72 1 SYSTEM CHART 2 NOMENCLATURE 3 HARDWARE INSTALLATION 16 22 4 DP2 TWAI...

Страница 4: ...Manual 6 Connect the power cord correctly and ensure that the grounding terminal of the power supply and wall outlet are properly connected If the equipment is not grounded earthed Olympus can no long...

Страница 5: ...ies for the operation errors and functional faults incurred due to the alteration of the environmental setup including BIOS change installation of other software or addition of hardware to the compute...

Страница 6: ...may be noticeable due to the difference in brightness between the transmitted sections over exposure and non transmitting section under exposure To reduce the flare set a lower exposure using the exp...

Страница 7: ...he movie recording time is limited according to the HDD free space 4 RAM If a RAM other than a PC2700 or greater dual channel DDR DDR2 RAM is used the full size live frame rate may drop 5 Monitor Us a...

Страница 8: ...l problem when the system is installed or run The space required for saving an image file in the HDD is slightly more than 4 MB with a 1360 x 1024 pixel 24 bit non compressed image and slightly more t...

Страница 9: ...ints or oil smudges wipe it with gauze slightly moistened with commercially available absolute alcohol Since the absolute alcohol is highly flammable it must be handled carefully To prevent fire ignit...

Страница 10: ...applicable For details please consult Olympus 1 1 DP72 For Desktop Computer DP2 BSW Basic Software Ver 2 1 or later DP2 TWAIN TWAIN Driver DP72 HDD to FDD power conversion cable SATA to HDD 4 pin powe...

Страница 11: ...when the DP72 PXU is combined DP72 Camera head AC adapter DP72 PXU PCIe Extension Unit PCIe extension box PCIe x1 cable ExpressCard Laptop computer SATA to HDD 4 pin power conversion adapter 1 2 DP72...

Страница 12: ...rantee any performance of the camera Connector P 14 C mount thread P 13 Tripod adapter mount screw P 13 PCIe Interface Board PCI standard half size low profile size compatible Trigger cable connector...

Страница 13: ...ion Unit For Laptop Computer PCIe Extension Box PCIe cable connector P 15 AC adapter connector P 15 PCIe x1 cable P 15 Vertical installation plates P 12 With 2 screws ExpressCard PCIe cable connector...

Страница 14: ...lation The SATA to HDD 4 pin power conversion adapter is sensitive to excessive force Please handle it with care 1 Open the cover of the computer 2 Remove the clamping screw of the slot cover of an un...

Страница 15: ...is attached on the back of the top cover Remove the tapes and the protective sheet 8 carefully so as not to displace the thermal conductive sheet Fig 5 7 Place the top cover in the original position a...

Страница 16: ...the U CMAD3 C mount adapter into the C mount thread on the bottom of the camera head Then loosen the clamping screw at the bottom of the U TV1X 2 camera adapter screw in the camera adapter into the C...

Страница 17: ...the interface cable into the connector on the camera head until it clicks and check that the interface cable is will not slip out 2 Push the connector on the other end of the interface cable into the...

Страница 18: ...up and insert ExpressCard 54 slot Hold the card with the surface marked INSERT facing up and insert it along the left edge of the slot 2 Check the orientation of the connector of the PCIe x1 cable an...

Страница 19: ...instruction manuals for your computer Before Installation Trademark Information Windows is a registered trademark of Microsoft Corporation All other brand and product names are trademarks or registere...

Страница 20: ...task to be used and click on the Next button 8 Enter the user information User Name Company Name and click on the Next button 9 The Choose Destination Location window appears If you want to change the...

Страница 21: ...stem and Maintenance Make sure not to click on Get started with Windows or Back up your computer 17 Click on System 14 Take out the CD ROM 15 To check that the installation is completed successfully c...

Страница 22: ...n the User Account Control dialog box appears click on the Continue button 20 If the driver has been properly installed MalAd Device should be displayed as shown below After confirming this click on t...

Страница 23: ...exe in the CD ROM drive folder and click on the OK button 6 Wait a while until the preparation for installation completes and the following window appears 9 The Choose Destination Location window appe...

Страница 24: ...lick on Control Panel 15 Click on Performance and Maintenance 11 The Start Copying Files window appears Click on the Next button 10 The Select Additional Tasks window appears Check the task to be used...

Страница 25: ...7 When the System Properties dialog box appears click on Hardware and then on Device Manager 18 If the driver has been properly installed MalAd Device should be displayed as shown below After confirmi...

Страница 26: ...mpatible application Brightfield Observation Fluorescence Observation Launch TWAIN compatible application Launch TWAIN compatible application Launch DP2 TWAIN Launch DP2 TWAIN Set up the microscope Se...

Страница 27: ...at the startup and used to record still images and movies For details refer to the DP2 TWAIN Online Manual 5 White balance adjustment 7 Black balance adjustment Exposure time setting 6 Image recording...

Страница 28: ...ith the DP2 BSW The trigger input is not compatible with random triggering Due to a time lag of 150 to 600 ms after the trigger input the exposure start time cannot be synchronized completely For deta...

Страница 29: ...close operations of the DP2 BSW The external trigger signal is output by starting still image recording when the shutter is closed The time lag from the trigger output to the start of exposure can be...

Страница 30: ...O speed ISO 200 400 800 1600 equivalent A D 12 bits Exposure control Exposure modes Auto SFL Auto Manual AE lock Available AE pause Available Exposure correction Correction range 2 0 EV Step 1 3 EV Me...

Страница 31: ...ntrast modes Low Standard High Linear Sharpness filter Low Standard High Color space sRGB Adobe RGB Focus indicator Contrast bar Line profile Interval recording Interval 1 sec to 24 h 59 m 59 sec Numb...

Страница 32: ...5 grams AC adapter 60 W x 110 D x 34 H mm excluding projections approx 250 grams Cord length Approx 1 0 meter PCIe x1 cable Approx 1 7 meter Operating environment These items can be set from the DP2 B...

Страница 33: ...face cable 8 The DP2 TWAIN or DP2 BSW is process ing recording Wait until the recording processing completes before starting recording of the next image In certain cases it may be required to press th...

Страница 34: ...s diaphragm a little The field iris diaphragm is not set prop erly Adjust the field iris diaphragm until the image circumscribes the field of view Lens components of the microscope are contaminated or...

Страница 35: ...If you do not want to retain the settings click on the No button 7 When the Maintenance Complete dialog box appears click on the Finish button Now the application is deleted 2 Windows R XP 1 Click on...

Страница 36: ...Table 1 Certified Cord A power supply cord should be certified by one of the agencies listed in Table 1 or comprised of cordage marked with an agency marking per Table 1 or marked per Table 2 The fit...

Страница 37: ...alter sterreichischer Verband f r Elektrotechnik VE Danmarks Elektriske Materialkontroll DEMKO National Standards Authority of Ireland NSAI Norges Elektriske Materiellkontroll NEMKO Asociacion Electro...

Страница 38: ...MEMO...

Страница 39: ......

Страница 40: ...ny 3500 Corporate Parkway P O Box 610 Center Valley PA 18034 0610 U S A One Corporate Drive Orangeburg NY 10962 U S A 491B River Valley Road 12 01 04 Valley Point Office Tower Singapore 248373 31 Gilb...

Отзывы: