Problem
Cause
Remedy
Page
1. The LED illumination does not
light.
The AC adapter or the power cord is
disconnected.
Connect the power cord to a power
outlet securely.
42
2. The brightness of observed field
of view is uneven and dark.
The objective is not engaged in the
light path properly.
Rotate the revolving nosepiece until it
clicks to engage the objective in the
light path.
—
The objective, eyepiece, condenser,
window lens or/and specimen are
dirty.
Clean them thoroughly.
4
The condenser top lens is not engaged
in the light path properly.
Push the top lens until it touches the
end.
—
The condenser turret position is not
correct.
Place the condenser turret properly
according to the observation method.
19
3. Dust or stains are visible in the
observed field of view.
The objective, eyepiece, condenser,
window lens or/and specimen are
dirty.
Clean them thoroughly.
4
4. The observed image glares.
The aperture diaphragm lever is
narrowed down too far.
Move the position of the aperture
diaphragm lever knob to the position
of the magnification same as the
magnification of the objective in use.
20
5. The observed image is blurred
in white or unclear.
The objective is not engaged in the
light path properly.
Rotate the revolving nosepiece until it
clicks to engage the objective in the
light path.
—
The objective, eyepiece, condenser,
window lens or/and specimen are
dirty.
Clean them thoroughly.
4
The immersion oil is not used with an
oil immersion objective.
Use the immersion oil.
21
The immersion oil contains bubbles.
Remove the air bubbles.
21
The specified immersion oil is not
used.
Use the immersion oil provided by
Olympus.
21
6. One-sided blur is found.
The image looks as if flowing.
The objective is not engaged in the
light path properly.
Rotate the revolving nosepiece until it
clicks to engage the objective in the
light path.
—
The specimen is not placed on the
stage properly.
Place the specimen on the stage
properly and secure it using the
specimen holder
10
7. The high magnification objective
collides with the specimen
before the specimen comes into
focus.
The specimen is placed upside
down.
Place the specimen with the cover
glass side facing up.
10
8
Troubleshooting
The performances of this microscope may not be fully delivered depending on how it is used, though that is not a failure.
If problems occur, please review the following list and take remedial action as needed.
If you cannot solve the problem after checking the entire list, please contact Olympus for assistance.