33
CX43
Problem
Cause
Remedy
Page
8. The tension of the coarse
focusing knob is too heavy.
The tension adjustment ring is
tightened too firmly.
Loosen the tension adjustment ring to
get proper tension.
14
9. The specimen is not in focus.
(The stage cannot be raised.)
The pre-focusing lever is positioned
too low.
Raise its position.
15
10. The stage descends by its own
weight or the focus is lost due to
slippage of the coarse focusing
knob.
The tension adjustment ring is
loosened too much.
Tighten the tension adjustment ring to
get proper tension.
14
11. The crossed Nicol is not set.
The analyzer is removed from the light
path.
Engage the analyzer in the light path.
28
12. The field of view is not matched
between two eyes.
The interpupillary distance is not
adjusted correctly.
Adjust it correctly.
16
The difference in diopter of two eyes
is not corrected properly.
Correct it properly.
17
Different eyepieces are used for right
side and left side.
Use the same eyepiece for right side
and left side.
39
13. When changing the objective
from lower magnification to
higher magnification, the
objective collides with the
specimen.
The specimen is placed upside
down.
Place the specimen with the cover
glass side facing up.
10
The cover glass is too thick.
Use a cover glass with thickness of
0.17 mm.
10
14. When changing the
magnification, the specimen is
defocused significantly.
The diopter is incorrect.
Adjust the diopter of eyepieces
correctly.
17
Repair request
If you cannot solve the problems even though taking actions described in "Troubleshooting", please contact Olympus
for assistance. Please provide us the following information at that time.
Product name and abbreviation (Ex.: Biological Microscope CX43RF)
Product number
Phenomena