According to procedures on page 38, attach the simple polarizing
intermediate attachment and the tube.
Insert the analyzer for simple polarization a into the slider insertion
opening of the simple polarizing intermediate attachment with the
display surface facing up until it touches the end, and tighten the
fixing knob b .
According to procedures on page 41, attach the polarizer for simple
polarization c to the filter holder (window lens) of the base portion
of the microscope frame.
Engage the 10X objective in the light path.
While looking through the eyepieces, rotate the polarizer for simple
polarization c with your hand to set to the status where the field of
view becomes the darkest (crossed Nicol).
1
2
333
4
5
1
Adjusting the crossed Nicol
a
b
c