DMTA069-01EN, Rev. D, March 2017
Chapter 7
68
7.2
Connector Reference
ChainSCANNERs sold after July 2013 come standard with the LEMO connector
which is compatible with the OmniScan MX2 and SX instruments. For use with a
different instrument, an optional adaptor is required (see Table 8 on page 68).
Figure 7-1 on page 69 and Table 9 on page 69 contain the pinout for the LEMO
connector used on the OmniScan MX2 model.
Table 7 Operating environment specifications
Parameter
Value
Operating temperature
5 °C to 50 °C (41 °F to 122 °F)
Storage temperature
‒
30 °C to 60 °C (
‒
22 °F to 140 °F)
Relative humidity (RH)
Max. 80 % RH noncondensing
Wet location
Yes
Altitude
Up to 2000 m
Outdoor use
Yes
Pollution level
1
IP rating
Waterproof (designed to meet IP67)
Table 8 Required encoder cable adaptor
Scanner
connector
Instrument
OmniScan MX
OmniScan MX2
OmniScan SX
TomoScan
FOCUS LT
LEMO (from July
2013 onward)
Omni-A-ADP27
[U8780329]
—
—
C1-LF-BXM-0.3M
[U8769010]
DE15 (Prior to
July 2013)
—
Omni-A2-ADP20
[U8775201]
Omni-A2-ADP20
[U8775201]
C1-DE15F-BXM-
0.30M [U8767107]
Содержание ChainSCANNER
Страница 5: ...DMTA069 01EN Rev D March 2017 Table of Contents v Index 87 ...
Страница 6: ...DMTA069 01EN Rev D March 2017 Table of Contents vi ...
Страница 8: ...DMTA069 01EN Rev D March 2017 List of Abbreviations viii ...
Страница 18: ...DMTA069 01EN Rev D March 2017 Important Information Please Read Before Use 10 ...
Страница 20: ...DMTA069 01EN Rev D March 2017 Introduction 12 ...
Страница 28: ...DMTA069 01EN Rev D March 2017 Chapter 1 20 ...
Страница 64: ...DMTA069 01EN Rev D March 2017 Chapter 5 56 ...
Страница 74: ...DMTA069 01EN Rev D March 2017 Chapter 6 66 ...
Страница 78: ...DMTA069 01EN Rev D March 2017 Chapter 7 70 ...
Страница 84: ...DMTA069 01EN Rev D March 2017 Appendix A 76 ...
Страница 94: ...DMTA069 01EN Rev D March 2017 List of Tables 86 ...