If you cannot improve the phenomena after taking the above remedies, please contact Olympus for assistance.
At that time, please tell them the following information as well.
· Product name and abbreviation (Example: stage with coaxial knobs on the bottom right U-SVRM)
· Product number
· Phenomena
Repair request
Phenomena
Cause
Remedy
Page
a) When the stage is touched by
the hand, the image moves
significantly.
The stage is not secured correctly.
Secure the stage firmly.
97
b) The stage does not move in the
Y-axis direction (front and back).
(U-SIC4R2/SIC4L2)
The Y-axis is locked.
Unlock the Y-axis.
21
c) The tension of the X-axis and
Y-axis knobs is too tight or too
loose.
(U-SVRM/SVLM)
The tension of the X-axis and Y-axis
adjustment knobs is too tight or too
loose.
Adjust the X-axis and Y-axis
adjustment knobs to the appropriate
tightness.
21
5-5 Stage
Phenomena
Cause
Remedy
Page
Number of beeps
CB
LIM /
HS
MIX
k) The stored brightness (LIM function) cannot be restored to the factory default setting.
None
/
–
–
The time for pressing the LIM switch and the SET switch on
the front of the microscope frame is short. (5 sec. or less)
Set the main switch to (ON) while pressing the LIM switch and
the SET switch, and wait 5 seconds while pressing the LIM
switch and the SET switch.
16, 30