Phenomena
Cause
Remedy
Page
Number of beeps
CB
LIM /
HS
MIX
e) When the button of the hand switch for motorized nosepiece (BX3M-HSRE) is pressed, the motorized nosepiece rotates, but it does
not reach the specified hole.
5 times
/
–
–
The setting of the dip switch is not correct.
Set the main switch to (OFF), and use the dip switch on the
side of the control box (BX3M-CB) to set the number of objective
mounting holes on the nosepiece.
16, 100
The motorized nosepiece did not reach the click position
within the specified time.
The motorized nosepiece is damaged.
Set the main switch to (OFF) and check that there is no
obstacle when the nosepiece rotates.
If the nosepiece contacts with the obstacle, remove the obstacle.
Then, set the main switch to (ON). If the same phenomena still
occurs, contact Olympus.
16
f) The MIX slider for reflected light observation (U-MIXR) cannot be controlled by operating the hand switch (BX3M-HS).
None
/
Turns OFF
The control box (BX3M-CB) is not connected.
Set the main switch to (OFF) and connect the control box and
the AC adapter of the control box again.
Then, set the main switch to (ON) and confirm that the beep is
heard once.
16, 100
, 117
The control box (BX3M-CB) is damaged.
If the indicator on the back of the controller does not light even
though the control box and the AC adapter of the control box
are connected again, and also if the beep sound is not heard
when the main switch is set to (ON), contact Olympus.
16, 100
, 117
None
/
Turns ON
The hand switch (BX3M-HS) is not connected.
Set the main switch to (OFF) and connect the hand switch
(BX3M-HS). If the indicator3 on the back of the control box does
not light even though the hand switch is connected, contact
Olympus.
16, 115
The hand switch (BX3M-HS) is damaged.
If the indicator3 on the back of the control box does not light
even though the hand switch (BX3M-HS) is connected, contact
Olympus.
115
None
/
Turns OFF
The MIX slider for reflected light observation (U-MIXR) is not
connected.
Connect the MIX slider for reflected light observation.
If the indicator2 on the back of the control box does not light
even though the MIX slider for reflected light observation is
connected, contact Olympus.
105
The MIX slider for reflected light observation (U-MIXR) is
damaged.
If the indicator2 on the back of the control box does not light
even though the MIX slider for reflected light observation is
connected, contact Olympus.
105
None
/
Turns ON
The cable for U-MIXR (U-MIXRCBL) or the extension cable
for U-MIXR (U-MIXRECBL) is disconnected.
Contact Olympus.
–