4-5 Reflected light infrared observation procedures
·
Combining the reflected polarizer slider for IR and the
rotatable analyzer slider for IR (U-POIR, U-AN360IR) cuts
the reflection to the sample surface to acquire the image
with the higher resolution.
·
Using the correction collar mechanism of the objective
reduces the aberration to acquire the bright image with the
high resolution.
NOTE
Start observation.
Brightfield/darkfield selector knob of the
reflected light illuminator
(page 17)
Coarse / Fine focusing knob of the
microscope frame
(page 24)
Insert the band-pass filters for IR
(U-BP1100IR, U-BP1200IR, U-25).
5
Turn ON the main switch.
1
Filter insertion slot of the reflected light
illuminator
(page 43)
Main switch
(page 16)
Light path selection lever of the trinocular
tube
(page 19)
Nosepiece
(page 23)
Coarse / Fine focusing knob of the
microscope frame
(page 24)
Select the brightfield observation.
3
Bring the sample top surface into focus
with the visible wavelength.
4
Change to the camera light path
6
Engage the objective with the
magnification to use in the light path.
7
Bring the sample into focus.
8
Transmitted/reflected light switch (page 16)
Select the reflected light illumination
2