![Olympus BX53M Скачать руководство пользователя страница 118](http://html1.mh-extra.com/html/olympus/bx53m/bx53m_instructions-manual_4020085118.webp)
112
15
Attaching the fluorescence mirror unit
1
2
The fluorescence mirror unit must be attached when the coded
universal reflected light illuminator (BX3M-URAS-S) is combined
with the microscope.
When attaching the fluorescence mirror unit, be sure to
engage the shutter in the light path for safety purpose.
Engage the shutter in the light path.
NOTE
1
Hold the both sides of the cover a on the front of the illuminator
lightly and pull it to the front side.
2
a
a
3
4
5
b
Turn the turret until the position number b of the fluorescence mirror
unit to be replaced comes to the front side, and stop it when the
clicking sound is heard.
Be careful not to allow your fingers being caught.
3
CAUTION
Hold the side of the fluorescence mirror unit to be replaced, and pull
it to the front side to remove it.
Insert the new fluorescence mirror unit in the position where the old
fluorescence mirror unit was removed until it touches the end.
4
5
Содержание BX53M
Страница 51: ...45 BX53M MEMO...
Страница 79: ...73 BX53M...