7
– Pre-installation Hardware Configuration
73
Chapter 7
| Pre-installation Hardware
Configuration
These steps only need to be followed if one or both of the 4-20 mA
inputs (
ANA1/ANA2
sensor inputs) is used.
This procedure must be performed by qualified, licensed
personnel. Since transmitters are factory configured, there is no
need to adjust these settings unless the configuration changes.
Since solder joints need to be created, this procedure must be
performed in a workshop with a non-explosive atmosphere.
The
OLCT 80
must be disconnected from power during the
soldering procedure.
Purpose
This procedure is used to configure the connections on the printed circuit board
for 2 auxiliary inputs (
In1
and/or
In2
)
depending on the type of sensor to be
connected (
4-20 mA
with 2, 3 or 4 wires).
Access the internal printed circuit board
Remove the display circuit board as follows:
■
Open the housing in a non-hazardous zone.
■
Remove the 4 screws used to secure the display circuit board. Remove the
circuit board. The flat connection cable and the lower printed circuit board
can remain in place.
■
The lower printed circuit board is now accessible.
Locate the solder pads
There are 3 solder pads (Figure 35) for each auxiliary input:
■
In 1
input: pads PPS1, PPS2 and PPS7 (A and B in the figure).
■
In 2
input: pads PPS3, PPS4 and PPS8 (A and B in the figure).
Содержание olct 80
Страница 1: ...1 User manual OLCT 80 Part Number NPO80GB Revision E 1 The Fixed Gas Detection Experts...
Страница 6: ...Table of Contents vi...
Страница 10: ...1 Overwiew 4...
Страница 18: ...2 Transmitter Overview 12...
Страница 22: ...3 Menus 16 Figure 15 the OLCT 80 s sub menus under AFF MES DATE TIME and PROG...
Страница 23: ...3 Menus 17 Figure 16 the OLCT 80 s sub menus under MAINT CALIBRA 4 20mA INFOS and TEST...
Страница 35: ...3 Menus 29 Figure 23 the ANA1 Configuration menu Press ESC repeatedly to return to the reading display...
Страница 39: ...3 Menus 33 Figure 25 the Alarm Relay Configuration menu Press ESC repeatedly to return to the reading display...
Страница 42: ...3 Menus 36 Figure 26 RELAY 1 menu Press ESC repeatedly to return to the reading display...
Страница 46: ...3 Menus 40 Figure 28 the Alarm Relay Configuration menu Press ESC repeatedly to return to the reading display...
Страница 58: ...3 Menus 52...
Страница 74: ...5 Wireless Version 68...
Страница 82: ...7 Pre installation Hardware Configuration 76...
Страница 84: ...8 Preventative Maintenance 78...
Страница 90: ...10 Accessories 84...
Страница 96: ...12 EU Declarations of Conformity 90 OLCT 80 without antenna...
Страница 97: ...12 EU Declarations of Conformity 91...
Страница 98: ...12 EU Declarations of Conformity 92 OLCT 80 with antenna...
Страница 99: ...12 EU Declarations of Conformity 93...
Страница 100: ...12 EU Declarations of Conformity 94 IR20 remote control of the OLCT 80...
Страница 108: ...13 Technical Specifications 102...
Страница 116: ...14 Special instructions for use in explosive environments and fonctional safety 110...
Страница 119: ...113...