Start-up
Preliminary inspection
Check the following:
■
That wiring was performed correctly.
■
That the detector housing is grounded.
■
That the braided shield of the connection cable is connected to the ground of
the central system.
■
That the device is securely mounted (screws, cable gland, cover screwed on
and locked).
Powering the detector
1. Block the central measuring controller or PLC to avoid accidentally triggering
any alarms during the procedure.
2. Power the detector.
3. Once the reading has stabilized, switch the central controller to normal
mode.
Stabilization time
After the device is mounted, it is important to allow the detector's temperature to
stabilize. Also, once the detector is powered, certain sensors require additional
pre-heating time If adjustments are made before the time indicated below has
passed, readings may be incorrect, which could put people and goods in
danger. Total wait time is summarized below:
■
Combustible sensor: 2 hours
■
Oxygen sensor: 1 hour
■
Electrochemical sensor: 1 hour, except for:
- NO (nitric oxide):
12 hours
- HCl (hydrochloric acid):
24 hours
- ETO (ethylene oxide):
36 hours
■
Semiconductor sensor: 4 hours
■
Infrared sensor (XPIR): 2 hours
Gas reading display
Normal display (no fault)
■
The display indicates the concentration measured,
the type of gas and the unit for the selected
channels (
Channel Configuration
menu, on page
Erreur ! Signet non défini.
).
■
The indicator light ( ) blinks.
Figure 54: display
under normal
operating conditions.
Содержание olct 80
Страница 1: ...1 User manual OLCT 80 Part Number NPO80GB Revision E 1 The Fixed Gas Detection Experts...
Страница 6: ...Table of Contents vi...
Страница 10: ...1 Overwiew 4...
Страница 18: ...2 Transmitter Overview 12...
Страница 22: ...3 Menus 16 Figure 15 the OLCT 80 s sub menus under AFF MES DATE TIME and PROG...
Страница 23: ...3 Menus 17 Figure 16 the OLCT 80 s sub menus under MAINT CALIBRA 4 20mA INFOS and TEST...
Страница 35: ...3 Menus 29 Figure 23 the ANA1 Configuration menu Press ESC repeatedly to return to the reading display...
Страница 39: ...3 Menus 33 Figure 25 the Alarm Relay Configuration menu Press ESC repeatedly to return to the reading display...
Страница 42: ...3 Menus 36 Figure 26 RELAY 1 menu Press ESC repeatedly to return to the reading display...
Страница 46: ...3 Menus 40 Figure 28 the Alarm Relay Configuration menu Press ESC repeatedly to return to the reading display...
Страница 58: ...3 Menus 52...
Страница 74: ...5 Wireless Version 68...
Страница 82: ...7 Pre installation Hardware Configuration 76...
Страница 84: ...8 Preventative Maintenance 78...
Страница 90: ...10 Accessories 84...
Страница 96: ...12 EU Declarations of Conformity 90 OLCT 80 without antenna...
Страница 97: ...12 EU Declarations of Conformity 91...
Страница 98: ...12 EU Declarations of Conformity 92 OLCT 80 with antenna...
Страница 99: ...12 EU Declarations of Conformity 93...
Страница 100: ...12 EU Declarations of Conformity 94 IR20 remote control of the OLCT 80...
Страница 108: ...13 Technical Specifications 102...
Страница 116: ...14 Special instructions for use in explosive environments and fonctional safety 110...
Страница 119: ...113...