64
BM 25/25W
Instruction Manual
Table No. 4
Chlorine
(Cl
2
)
Hydrochloric acid
(HCl)
Hydrogen cyanide
(HCN)
Sensor P/N
6313809
6313804
6313805
Standard range (1)
10
30
10
Measurement principle
Electrochemical
Electrochemical
Electrochemical
Display resolution(1)
0.1
0.1
0.1
Accuracy (2)
0.25
1
0.25
Repeatability (3)
2
2
2
Zero/Span ndrift (4)
0.5 / 5
0.5 / 5
0.5 / 5
Response time (5)
< 60s
< 80s
< 60s
Temperature (6)
-20°C to +40°C
-20°C to +40°C
-20°C to +40°C
Relative humidity and
pressure range (7)
■
10
– 90% RH
■
b
ar ± 20%
■
15
– 95% RH
■
1 bar ± 20%
■
15
– 95% RH
■
1 bar ± 20%
Lifetime (8)
30 months
24 months
24 months
Storage conditions and
maximum storage time
(9)
■
4°C
– 20°C
■
10
– 60% RH
■
1 bar ± 10%
■
2 months
■
4°C
– 20°C
■
10
– 60% RH
■
1 bar ± 10%
■
2 months
■
4°C
– 20°C
■
10
– 60% RH
■
1 bar ± 10%
■
2 months
Warm-up Delay (10)
Operational upon start up
Notes
■
Exposure to high levels of organic solvents can damage the sensors.
■
Exposure to gases at levels higher than the monitor’s range can
damage the device.
Recalibrate sensors if they go out of range.
Содержание BM25
Страница 1: ...BM 25 25W Area Gas Monitor Part Number NPB25EN Version L 2 User Manual ...
Страница 8: ...8 BM 25 25W Instruction Manual ...
Страница 10: ...10 BM 25 25W Instruction Manual Product Overview Figure 2 Product Overview ...
Страница 18: ...18 BM 25 25W Instruction Manual ...
Страница 34: ...34 BM 25 25W Instruction Manual ...
Страница 70: ...70 BM 25 25W Instruction Manual ...
Страница 76: ...76 BM 25 25W Instruction Manual ...
Страница 78: ...78 BM 25 25W Instruction Manual ...
Страница 79: ...10 Declaration of EC Conformity 79 ...