Alarm relay configuration and logic inputs
■
Set relay output and the logic input by
checking the different boxes (Figure 45).
Reference 1
Alarm relay output settings. In this example
all CO, O2 and LEL alarms will trigger the
relay output.
Reference 3
To set the BM 25/25W modes when logic
input is activated (set to 1).
-
Alarm relay
: internal relay output is
activated
-
Alarm 1
: BM 25/25W will turn in Alarm 1
mode
-
Alarm 2
: BM 25/25W will turn in Alarm 2
mode
-
Remote alarm
: BM 25/25W will turn in
Alarm Transfer mode
Reference 4
Alarm acknowledgment
-
Local acknowledgment
: alarms must be
acknowledged from the gas monitor
keyboard.
-
Remote acknowledgment
: when on a
network (wired or wireless), local alarms
can be acknowledged from another BM
25/25W.
■
Click
OK
to confirm.
Figure 45: alarm output relay and
logic input settings