OLCT 700/710 TP
Page 32 of 45
Rev. C.1
OLCT 700/710 TP Instruction Manual
Stability Fault
Probable Causes: Failed Sensor, empty or close to empty Cal Gas Cylinder, problems with cal gas and
delivery, or no Humidifying Tube used, failure to properly desiccant packet the sensor during extended
power-off periods.
•
Check Heater Power Setting (should be 235 +/- 5mW at 25ºC ambient temperature).
•
Verify use of span gas Humidifying Tube.
•
If using Splashguard with Integral Cal Port, must use Calibration Wind Guard or air movement can
compromise span gas delivery.
•
Check validity of span gas using H
2
S pull tube or other means (check MFG date on cal gas
cylinder).
•
Check for obstructions through stainless steel flame arrestor (including being wet, blocked, or
corroded).
•
Replace the plug-in H
2
S sensor.
Clearing Fault
Probable Causes: Failed Sensor, Cal Gas not removed at appropriate time or problems with cal gas and
delivery, presence of background H
2
S, or incorrect Heater Power Setting.
•
Must recover to < 5ppm in < 5 min after AutoSpan is complete.
•
Use bottled air (zero air) if there is a known continuous H
2
S background level.
•
Verify use of span gas Humidifying Tube.
•
Check validity of span gas using H
2
S pull tube or other means (check MFG date on cal cylinder).
•
Check for obstructions through stainless steel flame arrestor (including being wet, blocked, or
corroded).
•
Perform Heater Power Setting.
•
Replace the plug-in H
2
S sensor.
Poor Calibration Repeatability
Probable Causes: Failed Sensor, use of wrong Cal Gas or problems w/ cal gas and delivery, or Interference
Gases.
•
Check for adequate Sensor Life.
•
Check Heater Voltage Setting (should be 235±5mW at 25ºC).
•
Verify use of span gas Humidifying Tube.
•
Check validity of cal gas using H
2
S pull tube or other means (check MFG date on cal cylinder).
•
Check for obstructions through stainless flame arrestor (including being wet, blocked, or
corroded).
•
Evaluate area for presence of any contaminating gases as listed in Section 2.4.
•
Note the sensor’s serial # and report repetitive problems to Oldham’s Repair Department.
•
Replace plug-in H
2
S sensor.
Unstable Output/ Sudden spiking
Possible Causes: Unstable power supply, inadequate grounding, or inadequate RFI protection.
•
Verify Power source is stable.
•
Verify field wiring is properly shielded and grounded.
•
Contact Oldham to optimize shielding and grounding.
•
Add Oldham’s RFI Protection Circuit accessory if problem is proven RFI induced.
Содержание 700 TP Series
Страница 6: ...OLCT 700 710 TP vi OLCT 700 710 TP This page has been left blank intentionally...
Страница 20: ...OLCT 700 710 TP Page 14 of 45 Rev C 1 OLCT 700 710 TP Instruction Manual...
Страница 44: ...OLCT 700 710 TP Page 38 of 45 Rev C 1 OLCT 700 710 TP Instruction Manual 7 3 OLCT 700710 TP Drawings...
Страница 46: ...OLCT 700 710 TP Page 40 of 45 Rev C 1 OLCT 700 710 TP Instruction Manual...
Страница 47: ...OLCT 700 710 TP OLCT 700 710 TP Instruction Manual Rev C 1 Page 41 of 45 8 EU Declaration of conformity...
Страница 48: ...OLCT 700 710 TP Page 42 of 45 Rev C 1 OLCT 700 710 TP Instruction Manual...