OLCT 700/710 TP
Page 26 of 45
Rev. C.1
OLCT 700/710 TP Instruction Manual
Temperature Fault
If the OLCT 700/710 is currently reporting an ambient temperature that is outside of the –40°C to +75°C
range, a “Temperature Fault” is declared. A “Temperature Fault” will cause the “Fault Detected” message
to scroll once a minute on the display. The Modbus™ fault register bit for Temperature Fault will be set
and will not clear until the fault condition has been cleared. If a Temperature Fault occurs, the 4-20mA
signal remains operational.
AutoSpan Fault
If 180 days has elapsed since the last successful AutoSpan, an AutoSpan Fault will be generated. An
“AutoSpan Fault” will cause the “Fault Detected” message to scroll once a minute on the display. The
Modbus™ fault register bit for AutoSpan Reminder Fault will be set and will not clear until the fault
condition has been cleared. If an AutoSpan Reminder Fault occurs, the 4-20mA signal remains operational.
Refer to the Troubleshooting Guide, Section 0, for guidance on fault conditions If you cannot fix the fault
condition, email our Technical Service team at
Содержание 700 TP Series
Страница 6: ...OLCT 700 710 TP vi OLCT 700 710 TP This page has been left blank intentionally...
Страница 20: ...OLCT 700 710 TP Page 14 of 45 Rev C 1 OLCT 700 710 TP Instruction Manual...
Страница 44: ...OLCT 700 710 TP Page 38 of 45 Rev C 1 OLCT 700 710 TP Instruction Manual 7 3 OLCT 700710 TP Drawings...
Страница 46: ...OLCT 700 710 TP Page 40 of 45 Rev C 1 OLCT 700 710 TP Instruction Manual...
Страница 47: ...OLCT 700 710 TP OLCT 700 710 TP Instruction Manual Rev C 1 Page 41 of 45 8 EU Declaration of conformity...
Страница 48: ...OLCT 700 710 TP Page 42 of 45 Rev C 1 OLCT 700 710 TP Instruction Manual...