Regressions between OKRA Pro BGMS results and the YSI 2300 for the capillary whole
blood samples:
Precision Results for venous whole blood samples.
The study shows that the OKRA Pro glucose monitoring
system compares well with the laboratory method.
Precision
Linear regression
95% CI Slope
95% CI Intercept
R
2
N
y= 0.9886x - 0.3843
(0.978, 0.999)
(-2.610, 1.842)
0.9916
300
Blood Conc.
Level
N
44 mg/dL
83mg/dL
125 mg/dL
208 mg/dL
332mg/dL
grand mean
300
43
81
127
209
332
pooled
variance
300
2.4
4.4
11.7
27.9
76.3
pooled
SD
300
1.6
2.1
3.4
5.3
78.7
95% CI
(1.57, 1.63)
(2.04, 2.16)
(3.33, 3.47)
(5.18, 5.42)
(8.53, 8.87)
pooled CV
(%)
300
3.7
2.6
2.7
2.5
2.6
Precision Results for control solutions.
Control Solution Level
N
45 mg/dL
111 mg/dL
307mg/dL
grand mean
600
51
111
311
pooled variance
600
4.8
7.5
60.2
pooled SD
600
2.2
2.7
7.8
95% CI
600
(2.18, 2.22)
(2.26, 2.73)
(7.67, 7.93)
pooled CV (%)
600
4.3
2.4
2.5
Alternative Site Testing Accuracy
The OKRA Pro glucose monitoring system was evaluated for alternative site testing
accuracy by comparing blood glucose results of alternate sitesmeasured by patients
and health professionals using OKRA Pro glucose monitoring system, and the results
measured by the YSI 2300 Autoanalyzer, a laboratory instrument. Glucose levels were
measured on 150 fresh capillary specimens at three different clinical centers.