When the control solution is applied to the top edge of the test strip, you should
get results within theexpected range printed on the label of the test strip vial. If the
control solution test results fall outside the range, repeat the test. Results that fall
outside the range may be caused by:
- Error in performing the test.
- Failure to shake the control solution bottle well (must shake vigorously).
- Failure to discard the first drop of control solution.
- Expired or contaminated control solution.
- Test strip deterioration.
- Meter malfunction.
- Control solution that is too warm or cold.
IMPORTANT NOTE
If the control solution test results continually fall outside the range printed on the vial, the OKRA
Pro glucose monitoring system may not be functioning properly. DO NOT use the system to test
your blood until you receive a control solution test result that falls within the correct range. If you
continue to have problems, contact your local representative immediately.
The OKRA Pro test strips provide accurate results when the following constraints
are observed: Inaccurate results may occur in severely hypotensive individuals or
patients in shock. Inaccurate low results may occur for individuals experiencing a
hyperglycemic-hyperosmolar state, with or without ketosis. Critically ill patients
should not be tested with this meter.
- Use only the OKRA Pro test strips with the OKRA Pro meter.
- Use fresh capillary whole blood only.
- The test strips are for single use only. Do not reuse.
- Dehydration may cause higher test results.
- Inaccurate results may occur when in shock, hypotensive, hyperglycemic, or
hyperosmolar state and with or without ketosis.
- OKRA Pro test strips used above altitudes of 10,000 feet will have an effect on test
results.
Limitations of System