Cleaning and disinfecting procedures
1. Wash your hands before
performing cleaning and
disinfecting procedures.
2. Remove the wipe from the container.
3. Make sure the meter is turned off.
4. Wipe the outside of the meter carefully
(3 passes horizontally and 3 passes vertically).
* To clean the meter, wipe it with a CaviWipes
TM
towelette.
* To disinfect the meter, wipe it with an other CaviWipes
TM
tow-
elette and leave it wet for 2 minutes.
NOTE
1. The meter should be cleaned prior to each disinfection.
(Two CaviWipes
TM
disinfecting towelettes in total.)
2. To achieve disinfection of the stated efficacy kill claim, the meter
should remain visibly wet for 2 minutes right after disinfecting
treatment.
5. Dispose of the used wipes.
6. Wash your hands thoroughly with soap and water.
The meter must be cleaned and disinfected a minimum of once per week.
1. Wash your hands before performing cleaning and disinfecting procedures.
2. Remove the wipe from the container.
3. Make sure the used lancet is removed prior to cleaning.
4. Wipe the outside of the reusable lancing device carefully (3 passes horizontal and
3 passes vertical).
* To clean the reusable lancing device, wipe it with a CaviWipes
TM
towelette.
Reusable lancing device
CAUTION
Take extreme care that the liquid does not soak into the data ports. Do not use any type of spray.
CAUTION
If the meter is being operated by a second person who is providing testing assistance to the user,
the meter and lancing device should be decontaminated prior to use by the second person.