57
13
CO2-O2-MODULE Calibration
Press on
Settings
icon
to access to the
Settings
subpage (see Figure 75 a). Press on
Calibration
icon
(see Figure 75 b) and select
CO2/O2
icon
(see Figure 75 c). The
CO2/O2 Calibration
page appears as
in Figure 76).
b
c
a
Figure 75. How to set CO2/O2 Calibration (a
–
b
–
c)
Figure 76. CO2/O2 Calibration page.
CO2/O2 Calibration
page contains five subpages:
•
View Current
: to view factory calibration or CO2/O2 User Calibration data. See paragraph 13.1.
•
Span Calibr.
: to run CO2/O2 User Calibration using certified span gas. See paragraph 13.2.
Note
►
Span gas is a gas mixture of precisely defined composition used as reference gas in the calibration of
gas meters. Calibration is an important issue for all analytical instruments and is done by exposing the sensor