to a gas of known concentration used as a comparative standard. The calibration procedure detects and
deletes the reading error.
•
Factory
: to cancel the last CO2/O2 User Calibration and restore the factory values. See paragraph
•
CO2 Zero Reset
: to correct the zero of the CO2 sensor. See paragraph 13.4.
•
Offset: to
set CO2 and/or O2 offset using an external gas meter. See paragraph 13.5.
13.1
View Current
LEO is equipped with a factory calibrated CO2-O2-MODULE using certified gas cylinders. The factory
calibration certificate is included in electronic format and can be seen in the
CO2/O2 Calibration
page. Press the
View Current
icon
(see Figure 77 a) to view the factory calibration or CO2/O2 User Calibration data. If no
CO2/O2 User Calibration has been made, there are only the
CO2
and/or O2
Factory Calibration
subpages, as in the
example shown in Figure 77 b and c.
a
b
c
Figure 77. CO2/O2 View Calibration (a
–
b
–
c).
The recommended CO2-O2-MODULE calibration interval is six months. However, Okolab suggests to
verify LEO’s reading each month or before each important measurement by employing a certified span gas
. To
achieve the best accuracy, perform the gas measurement immediately after the CO2/O2 User Calibration.
13.2
Span Calibration
To improve LEO reading accuracy, it is possible to perform CO2/O2 User Calibration using span gases.
If you want to achieve the best accuracy in the sensors’ measurement range 0 –
10%, Okolab suggests to
use a certified span gas cylinder with composition similar to the gas used during your working standard conditions.
In this case, after the CO2/O2 User Calibration you can achieve an accuracy of ±0.1%.
If you want to achieve the best accuracy in the sensors’
full measurement range, Okolab suggests to use up
to two span gases:
•
Certified span gas cylinder with composition similar to the gas used during the standard working
conditions
Содержание LEO
Страница 1: ...LEO Handheld Analyzer for IVF Applications Manual IST1446_REV05 SV 3 1 64 0 ...
Страница 2: ...This page was left blank ...
Страница 16: ...12 Figure 13 Pump ...
Страница 84: ...80 Figure 110 T1 Calibration Factory Reset a b b a ...
Страница 88: ...84 f e c a b d 2 1 1 Figure 114 CO2 O2 MODULE insertion ...
Страница 92: ...88 f e c a b d 2 1 1 Figure 118 CO2 O2 MODULE insertion after pump replacement ...