34
10
Enclosure standard components
H201-T-UNIT-BL includes the enclosure (cage) surrounding the microscope (
purchased separately
). The
enclosure depends on the user’s microscope therefore it can
be different for each microscope. There are some
parts of the enclosure that are, however, the same for all. The standard parts are shown in the following images.
A.
Humidity Module (included in HM-VF) positioning on the enclosure top panel
. You can insert this
part of the HM-VF also at the end of the assembly of the whole cage. See Figure 53.
Figure 53.HM-VF inserting in the appropriate slot
Tip
►
The Humidity Module allows to humidifying the mixed gas coming from the Gas Mixer by a permeable
water membrane. The dry gas flows through this permeable membrane immersed in water before arriving
to the chamber incubator. The Humidity Module has to be filled with distilled water until to cover entirely
the tubing into the reservoir, this operation can be done before inserting it into its housing or after the
insertion, you have to lift the black plug only (see Figure 54).
Fill the Humidity Module
with distilled water
Permeable Water
Membrane
Figure 54. How to fill the Humidity Module.
Note
►
This particular component is not present if you have purchased HM-CTIVE as Humidity Module. In
this case this bore is covered by a plug while the system configuration becomes the one shown in
B.
Water Trap (included in HM-VF) positioning inside the enclosure
. You can insert this part of the HM-
VF or before mounting the frontal panel or after have mounted the frontal panel simply opening the slide
door on the left panel of the enclosure. See Figure 55.