16
Check that no error message will be displayed on the Chiller’s Control Panel (see Chiller’s User
Manual, pag. 15).
2.
Set the Chiller’s pump pressure using the bypass adjustment wheel placed on the rear panel of the
Chiller (see Figure 16).
Figure 16
. Chiller’s bypass adjustment wheel.
To increase the pump pressure, turn the bypass adjustment wheel clockwise very slowly until a
pressure value of 1 barg is reached. To check the pressure value, monitor the manometer placed on
the front panel of the Chiller (see Figure 17).
Figure 17
. Chiller’s Manometer.
Set the Chiller’s pump pressure at 1 barg. Do not exceed 1.
3 barg.
Note
►
Make sure that the valves along the Water Tubes are opened.
3.
Set the lower limit value for the temperature of the Chiller at 2
°C (see the Chiller’s User Manual pag.
39).
4.
Switch the T-Unit on using the power button on OKO-TOUCH. A small green LED light will blink on
the front of the T-Unit. This means that the system is initializing.