12
T
ECHNICAL
S
PECIFICATIONS
T-Unit Technical Specifications
Temperature
Temperature Range: from 10/15°C BELOW Room temperature (depending on chamber model)
to 60°C
Sample Temperature range: 3°C above ambient temperature to 50°C
Step size: 0.1°C
Accuracy:0.2°C in Sample Mode and in Cage Mode if room temperature remains within ± 1°C.
Regime temperature time
Depends on the control mode, about 30 min in chamber mode feedback
Operating Temperature
0°C ~ +55°C
Storage Temperature
-5°C ~ +60°C
Operating Humidity
0-70%
Power Consumption
24 V DC 18 W max
Weight
1.55 Kg
Water Bath Supply Voltage
115V 60 Hz /230V 50 Hz
Water Bath Power Consumption
2.2 kW.
Water Bath Dimensions [LxWxH]
180x350x546