Usually the Knob is positioned in configuration 2, called medium position. If water doesn’t circulate in the
chamber, causing thermal uniformity absence please check the Knob position otherwise set it in position 1
The level of circulation pump is automatically set at the standard value 2. This value 2 ensures good
circulation and thermal uniformity in the incubation chamber. If you want to VERIFY or modify this value,
please follow the next scheme.
Tip
►
If the recirculation is impeded or reduced, the Stage Incubator will not be evenly warmed up. Avoid
narrowing, throttling, kinking of tubing, and formation of air bubbles to ensure system performances
.