6700-0235-000 (Rev.1) 09/2017

Specifications

Height = Approximately 4 inches  

Width = Approximately 2 inches

Vacuum: 

 

 

 

 

 

0 to 650 mm Hg

Operating Temperature Range:   

39.2°F to 120.2°F  (4°C to 49°C)

Storage Temperature Range: 

-0.4°F to 159.8°F  (-18°C to 71°C)

Operating and Storage Relative Humidity: 

5 to 95%

  

7700-0235-000 (Rev. 1) 2017-09

Product Setup:

CAUTION

 This device must be in the vertical position to function.

 Before each use, inspect the device for any damage; DO NOT USE if damaged or visibly contaminated. 

 Do not overtighten the Sure-Trap to vacuum regulator.

 Conduct visual pre-use check to ensure the deflector shield is attached to the cap.

Attach the Sure-Trap to inlet of Vacuum Regulator and attach 

suction tubing from the barb fitting to the vacuum port of 

collection canister as shown.

Cleaning

The exterior of the device may be cleaned with mild cleaning of disinfecting solution.

Disposal

WARNING

 After patient use, suction equipment may be contaminated.

If liquid is present in the device, remove from service and discard immediately.

⚠ 

Dispose of trap in accordance with the local regulations using biohazard precautions.