8700-0006-000 (Rev. 12) 08/2020

3/Description and Specifications

3.1 Description

The PTS-CVR is a lightweight, compact unit used throughout the hospital primarily for pharyngeal/tracheal suctioning 

(airway management). Various models provide regulated or full-line vacuum for hospital suction procedures.
There are several models of the PTS-CVR. All models contain a vacuum gauge which indicates suction supplied by 

the regulator. Each has a positive pressure safety relief valve to prevent pressurization by either failed injector vacuum 

(Venturi

®

) units or inadvertent cross connection to pressurized gasses. In addition, the Low PTS-CVR models include a 

vacuum relief valve to limit maximum suction.
Some models operate in a regulated or non-regulated  MAX  mode. Others operate only in the regulated mode.

In the non-regulated  MAX  mode, the vacuum source is connected directly to the fitting/patient port. The regulator module 

is bypassed and full-line vacuum is provided.
In the regulated mode, the vacuum source is connected through the regulator module which functions as an automatic 

valve. Turning the suction control knob adjusts the position of the regulator module and allows selection of a 

predetermined level of suction when set according to instructions.

During use, as the flow requirement increases, the valve automatically opens to maintain suction at the preset level. 

When the flow requirement decreases, the valve automatically closes to maintain suction at the preset level. The same 

mechanism compensates for changes in supply vacuum and automatically maintains the pre-set suction level when set 

according to instructions.

1.  Suction Control Knob - Allows easy adjustment of suction to the patient.
2.  Mode Selector Switch - Allows quick and easy mode changes.

a.  | (ON) - Suction can be adjusted with the suction control knob.
b.  O (OFF) - No suction is supplied to the patient.
c.  MAX  - Maximum full-line vacuum is supplied to the  patient.

3.  Vacuum Gauge - The suction level to the patient is displayed during use.

Mode 

Selector 

Switch

MAX

®

Vacuum 

Gauge

Adapter/Probe 

Port

Fitting/Patient Port

Suction 

Control 

Knob

Содержание Push-To-Set PTS-CVR

Страница 1: ...90 100 100 200 300 400 500 600 700 kPa mm Hg L L M M H H F F MEDICAL VACUUM ANSI Analog Adult Standard 1224 Adult High 1226 Digital Adult Standard 1324 Adult High 1326 ANSI Analog Adult Standard 1225...

Страница 2: ...surance Department The user of this Product shall have the sole responsibility for any malfunction which results from improper use faulty maintenance improper repair damage or alteration by anyone oth...

Страница 3: ...c Unit 13 6 3 2 2 Neonatal Unit 13 6 4 Regulator Module 13 6 4 1 Disassembly 14 6 4 2 Assembly 14 6 5 Digital Gauge 14 6 5 1 Removal of Batteries 14 6 5 2 Inserting Batteries 14 7 Service Checkout Pro...

Страница 4: ...ructions for Use Serial Number Manufacturer Abbreviations Used In This Manual CCW Counter clockwise decrease CW Clockwise increase CVR Continuous Vacuum Regulator MAX Full Line Vacuum inHg Inches of m...

Страница 5: ...ter will help prevent this and extend the life of suction equipment 1 3 CAUTION Do not use any Loctite products or any products which contain Methacrylate Ester as an active ingredient to seal the thr...

Страница 6: ...00 kPa mm Hg Analog 1230 Digital 1337 High Two Mode Continuous Adult L L M M H H F F MEDICAL VACUUM Analog 1330 Digital 1430 High Two Mode Continuous Adult mm Hg 100 200 500 700 300 400 600 MAX L L M...

Страница 7: ...uum is provided In the regulated mode the vacuum source is connected through the regulator module which functions as an automatic valve Turning the suction control knob adjusts the position of the reg...

Страница 8: ...l scale 1 Full scale High 0 760 mmHg 0 101 3 KPA 5 Full scale 1 Full scale Pediatric 0 160 mmHg 0 21 3 KPA 5 Full scale 1 Full scale Neonatal 0 160 mmHg 0 21 3 KPA 0 100 mmHg 0 13 3 KPA 5 Full scale 1...

Страница 9: ...ine system Failure to clean and disinfect the wall outlet and pipeline system may result in damage to this equipment 4 1 1 Routine Exterior Cleaning Routine cleaning of the regulator is recommended as...

Страница 10: ...ated products DO NOT SEND UNIT BACK TO THE MANUFACTURER Gauge doesn t respond to changes in suction Gauge assembly is not properly aligned Ensure gauge assembly 6 is properly aligned Missing Damaged O...

Страница 11: ...e assembly Replace gauge assembly 6 Whistling buzzing noise from the unit Need to apply grease on stem on regulator module Apply grease on stem 31 Gauge responds slowly to changes in suction failed bl...

Страница 12: ...est the gauge on its face Note See Section 9 1 Illustrated Parts 1 Remove the four cover screws from the back of the regulator 2 With regulator facing up carefully pull the cover assembly off the back...

Страница 13: ...with a drop of removable thread locker such as Loctite 242 Repeat step 5 to verify the vacuum relief valve setting 8 Re attach the cover see Section 6 2 2 Assembly WARNING Excess Loctite may seal the...

Страница 14: ...Place the O ring on the housing 5 Fold the diaphragm and place it on the piston head 6 Place the cap on the housing and press until it snaps into place 6 5 Digital Gauge Note Both batteries are positi...

Страница 15: ...ith an accuracy of 1 of full range at 22 C Note All gauge needles should come to rest within the zero range bracket or return to the stop pin when no suction is being supplied Gauges which do not comp...

Страница 16: ...1 mmHg 0 1 kPa for the NEO model and 8 mmHg 1 0 kPa for the high model 7 5 Maximum Vacuum Test PED NEO C A S S Models ONLY 1 Occlude the fitting patient port 2 Rotate the suction control knob fully c...

Страница 17: ...ting patient and adapter probe ports with minimum inner diameters of 0 140 The unit is either directly connected to the vacuum source or connected with a tube no more than 4 in length with a minimum i...

Страница 18: ...nvironmental conditions Replace lubricate and repair as needed 8 3 Repair Policy WARNING To reduce transportation personnel and or service personnel exposure to hazardous contamination clean and disin...

Страница 19: ...nd Fittings CAUTION Do not use any Loctite products or any products which contain Methacrylate Ester as an active ingredient to seal the threads on the adapters probes and fittings All Ohio Medical re...

Страница 20: ...acuum Relief Assembly 1 36 8700 0024 500 Quad Ring 006 Nitrile 1 18 See table below O Ring 2 014 Nitrile Buna N 1 8700 0006 000 PTS CVR Service Manual 1 19 6700 0115 400 Relief Valve Positive Pressure...

Страница 21: ...21 8700 0006 000 Rev 12 08 2020 9 Ordering Information 23 15 14 13 11 1 5 4X 7 4X 6 5 4 7 7 25 2 26 3 27 1 12 16 10 8 5 18 19 20 21 22 17 24 32 31 30 29 28 35 36 34 33...

Страница 22: ...hobic Filter 1 8 NPT x Tubing nipple 6730 0580 800 3 Pack 6730 0581 800 10 Pack 6730 0582 800 50 Pack Tubing 6700 0005 300 Reusable Overflow Safety Traps Description Part Number Overflow Safety Trap w...

Страница 23: ...C 61000 4 2 8kV Contact 2 kV 4 kV 8 kV and 15 kV Air 8kV Contact 2 kV 4 kV 8 kV and 15 kV Air Floors should be wood concrete or ceramic tile If the floors are covered with synthetic material the relat...

Страница 24: ...unity Test Level V m 385 380 390 TETRA 400 Pulse modulation 18 Hz 1 8 0 3 27 450 430 470 GMRS 460 FRS460 FM 5 kHz deviation 1 kHz sine 2 0 3 28 710 704 787 LTE Band 13 17 Pulse modulation 217 Hz 0 2 0...

Страница 25: ...x is a registered trademark of S C Johnson Son Inc Kleenaseptic is a registered trademark of Metrex Research Corporation Betco is a registered trademark of Betco Corporation Cavicide is a registered t...

Отзывы: