10/Electromagnetic Compatibility Declarations for PTS Digital

Guidance and Manufacturer's Declaration - Electromagnetic Emissions

The PTS Digital is intended for use in the electromagnetic environment specified below.  The user of the PTS 

Digital should assure that it is used in such an environment.

Emissions Test Compliance

Electromagnetic Environment - Guidance

RF emissions CISPR 

11

Group 1

The PTS Digital uses RF energy only for its internal functions.  

Therefore, its RF emissions are low and are not likely to cause any 

interference in nearby  

electronic equipment

RF Emissions CISPR 

11

Class A

The PTS Digital is suitable for use in all establishments other than 

domestic establishments and those directly connected to the public 

low-voltage power supply network that supplies buildings used for  

domestic purposes.

Guidance and Manufacturer's Declaration - Electromagnetic Immunity

The PTS Digital is intended for use in the electromagnetic environment specified below.  The user of the PTS 

Digital should assure that it is used in such an environment.

Immunity Test

IEC 60601-1-

2 Test Level Compliance Level

Electromagnetic Environment - 

Guidance

Electrostatic  

Discharge (ESD) IEC 

61000-4-2

±8kV  Contact                 

±2 kV, ±4 kV, ±8 

kV, and ±15 kV 

Air

±8kV Contact, ±2 kV, 

±4 kV, ±8 kV, and ±15 

kV Air

Floors should be wood, concrete, or 

ceramic tile.  If the floors are covered with 

synthetic material, the relative humidity 

should be at least 30%.

Radiated RF IEC 

61000-4-3

3 V/m 80 MHz 

to 2.7 GHz 80% 

AM at 1 kHz

3 V/m 80 MHz to 2.7 

GHz 80% AM at 1 kHz

Field strengths from fixed RF 

transmitters, as determined by an 

electromagnetic site survey*, should be 

less than the compliance level in each 

frequency range.  Interference may occur 

in the vicinity of equipment marked with 

this symbol:

Note: These guidelines may not apply in all situations.  Electromagnetic propagation is affected by absorption and reflection from 

structures, objects and people.

*Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, 

amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.  To assess the 

electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.  If the mea

-

sured filed strength in the location in which the PTS Digital is used exceeds the applicable RF compliance level above, the PTS 

Digital should be observed to verify normal operation.  If abnormal performance is observed, additional measures may be neces-

sary, such as reorienting or relocating the PTS Digital.   

 

10.1

Содержание Push-To-Set PTS-CVR

Страница 1: ...90 100 100 200 300 400 500 600 700 kPa mm Hg L L M M H H F F MEDICAL VACUUM ANSI Analog Adult Standard 1224 Adult High 1226 Digital Adult Standard 1324 Adult High 1326 ANSI Analog Adult Standard 1225...

Страница 2: ...surance Department The user of this Product shall have the sole responsibility for any malfunction which results from improper use faulty maintenance improper repair damage or alteration by anyone oth...

Страница 3: ...c Unit 13 6 3 2 2 Neonatal Unit 13 6 4 Regulator Module 13 6 4 1 Disassembly 14 6 4 2 Assembly 14 6 5 Digital Gauge 14 6 5 1 Removal of Batteries 14 6 5 2 Inserting Batteries 14 7 Service Checkout Pro...

Страница 4: ...ructions for Use Serial Number Manufacturer Abbreviations Used In This Manual CCW Counter clockwise decrease CW Clockwise increase CVR Continuous Vacuum Regulator MAX Full Line Vacuum inHg Inches of m...

Страница 5: ...ter will help prevent this and extend the life of suction equipment 1 3 CAUTION Do not use any Loctite products or any products which contain Methacrylate Ester as an active ingredient to seal the thr...

Страница 6: ...00 kPa mm Hg Analog 1230 Digital 1337 High Two Mode Continuous Adult L L M M H H F F MEDICAL VACUUM Analog 1330 Digital 1430 High Two Mode Continuous Adult mm Hg 100 200 500 700 300 400 600 MAX L L M...

Страница 7: ...uum is provided In the regulated mode the vacuum source is connected through the regulator module which functions as an automatic valve Turning the suction control knob adjusts the position of the reg...

Страница 8: ...l scale 1 Full scale High 0 760 mmHg 0 101 3 KPA 5 Full scale 1 Full scale Pediatric 0 160 mmHg 0 21 3 KPA 5 Full scale 1 Full scale Neonatal 0 160 mmHg 0 21 3 KPA 0 100 mmHg 0 13 3 KPA 5 Full scale 1...

Страница 9: ...ine system Failure to clean and disinfect the wall outlet and pipeline system may result in damage to this equipment 4 1 1 Routine Exterior Cleaning Routine cleaning of the regulator is recommended as...

Страница 10: ...ated products DO NOT SEND UNIT BACK TO THE MANUFACTURER Gauge doesn t respond to changes in suction Gauge assembly is not properly aligned Ensure gauge assembly 6 is properly aligned Missing Damaged O...

Страница 11: ...e assembly Replace gauge assembly 6 Whistling buzzing noise from the unit Need to apply grease on stem on regulator module Apply grease on stem 31 Gauge responds slowly to changes in suction failed bl...

Страница 12: ...est the gauge on its face Note See Section 9 1 Illustrated Parts 1 Remove the four cover screws from the back of the regulator 2 With regulator facing up carefully pull the cover assembly off the back...

Страница 13: ...with a drop of removable thread locker such as Loctite 242 Repeat step 5 to verify the vacuum relief valve setting 8 Re attach the cover see Section 6 2 2 Assembly WARNING Excess Loctite may seal the...

Страница 14: ...Place the O ring on the housing 5 Fold the diaphragm and place it on the piston head 6 Place the cap on the housing and press until it snaps into place 6 5 Digital Gauge Note Both batteries are positi...

Страница 15: ...ith an accuracy of 1 of full range at 22 C Note All gauge needles should come to rest within the zero range bracket or return to the stop pin when no suction is being supplied Gauges which do not comp...

Страница 16: ...1 mmHg 0 1 kPa for the NEO model and 8 mmHg 1 0 kPa for the high model 7 5 Maximum Vacuum Test PED NEO C A S S Models ONLY 1 Occlude the fitting patient port 2 Rotate the suction control knob fully c...

Страница 17: ...ting patient and adapter probe ports with minimum inner diameters of 0 140 The unit is either directly connected to the vacuum source or connected with a tube no more than 4 in length with a minimum i...

Страница 18: ...nvironmental conditions Replace lubricate and repair as needed 8 3 Repair Policy WARNING To reduce transportation personnel and or service personnel exposure to hazardous contamination clean and disin...

Страница 19: ...nd Fittings CAUTION Do not use any Loctite products or any products which contain Methacrylate Ester as an active ingredient to seal the threads on the adapters probes and fittings All Ohio Medical re...

Страница 20: ...acuum Relief Assembly 1 36 8700 0024 500 Quad Ring 006 Nitrile 1 18 See table below O Ring 2 014 Nitrile Buna N 1 8700 0006 000 PTS CVR Service Manual 1 19 6700 0115 400 Relief Valve Positive Pressure...

Страница 21: ...21 8700 0006 000 Rev 12 08 2020 9 Ordering Information 23 15 14 13 11 1 5 4X 7 4X 6 5 4 7 7 25 2 26 3 27 1 12 16 10 8 5 18 19 20 21 22 17 24 32 31 30 29 28 35 36 34 33...

Страница 22: ...hobic Filter 1 8 NPT x Tubing nipple 6730 0580 800 3 Pack 6730 0581 800 10 Pack 6730 0582 800 50 Pack Tubing 6700 0005 300 Reusable Overflow Safety Traps Description Part Number Overflow Safety Trap w...

Страница 23: ...C 61000 4 2 8kV Contact 2 kV 4 kV 8 kV and 15 kV Air 8kV Contact 2 kV 4 kV 8 kV and 15 kV Air Floors should be wood concrete or ceramic tile If the floors are covered with synthetic material the relat...

Страница 24: ...unity Test Level V m 385 380 390 TETRA 400 Pulse modulation 18 Hz 1 8 0 3 27 450 430 470 GMRS 460 FRS460 FM 5 kHz deviation 1 kHz sine 2 0 3 28 710 704 787 LTE Band 13 17 Pulse modulation 217 Hz 0 2 0...

Страница 25: ...x is a registered trademark of S C Johnson Son Inc Kleenaseptic is a registered trademark of Metrex Research Corporation Betco is a registered trademark of Betco Corporation Cavicide is a registered t...

Отзывы: