Ohio Medical Push-To-Set PTS-CVR Скачать руководство пользователя страница 18

8700-0006-000 (Rev. 12) 08/2020

18

8/Maintenance

8.1 General Maintenance of Suction Equipment

WARNING: The Pre-use Checkout Procedure must be performed before using this equipment on each patient.  

If the regulator fails any part of the Pre-use Checkout Procedure, it must be removed from service and 

repaired by qualified service personnel.

WARNING:  If the unit has been contaminated, clean and disinfect all suction equipment before disassembly.

Protection of the vacuum piping system is as important as maintenance of the suction equipment. The use of collection 

canisters with reliable shut-off valves, overflow safety trap assemblies and disposable suction filters will protect the 

regulator, wall outlet, and piping system. 
Routine maintenance and inspection are important to the performance of suction equipment. The following is a 

recommended list for care of suction equipment after each patient use. 
1. 

Keep connecting tubing, fittings and adapters/probes clean.

2.  Wipe all exterior surfaces with a solution of water and mild detergent or approved cleaning solutions.
3.  Perform a careful visual inspection of the vacuum regulator.

Item

Comments

Perform Service Checkout Procedure

If the regulator does not pass, refer to troubleshooting (Section 5 

Troubleshooting). Repair as needed

Check elastomeric parts, O-rings, gaskets, 

diaphragms, internal filters

Cleaning, lubrication and replacement interval depends on hours of 

usage and environmental conditions.  Replace, lubricate, and repair  

as needed

8.3 Repair Policy

WARNING: 

To reduce transportation personnel and/or service personnel exposure to hazardous contamination, clean 

and disinfect all suction equipment before shipping for service.

CAUTION: 

Do not steam autoclave or liquid sterilize the PTS-CVR. Severe impairment to the operation of the regulator 

will result. 

CAUTION: 

Only competent individuals trained in the repair of this equipment should attempt service.

Do not use malfunctioning equipment. Make all necessary repairs. Have the equipment repaired by qualified service 

personnel or by Ohio Medical. After repair, perform the Service Checkout Procedure to ensure that the product is 

functioning properly, and complies with the manufacturer’s published specifications.

4. 

Check that the high flow disposable suction filter is 

clean and in good condition.

5.  Check that all tubing is in good condition and 

connected securely to the correct ports.

6. 

Check the floats in the Overflow Protection Device 

and collection canister for correct operation.

7.  Perform the Pre-use Checkout Procedure.

8.2 Recommended Maintenance

We recommend the following to determine maintenance appropriate for each facility:
1.  Periodically inspect the overall condition of the vacuum regulator. Test gauge accuracy (Section 7.3 Gauge Test) and 

perform the pre-use checkout procedure (See PTS-CVR Instructions for Use for the Pre-use Checkout Procedure). If 

the regulator does not pass, refer to troubleshooting (Section 5 Troubleshooting).

2.  Determine maintenance based on data from your periodic inspections. Follow the guidelines below.

Содержание Push-To-Set PTS-CVR

Страница 1: ...90 100 100 200 300 400 500 600 700 kPa mm Hg L L M M H H F F MEDICAL VACUUM ANSI Analog Adult Standard 1224 Adult High 1226 Digital Adult Standard 1324 Adult High 1326 ANSI Analog Adult Standard 1225...

Страница 2: ...surance Department The user of this Product shall have the sole responsibility for any malfunction which results from improper use faulty maintenance improper repair damage or alteration by anyone oth...

Страница 3: ...c Unit 13 6 3 2 2 Neonatal Unit 13 6 4 Regulator Module 13 6 4 1 Disassembly 14 6 4 2 Assembly 14 6 5 Digital Gauge 14 6 5 1 Removal of Batteries 14 6 5 2 Inserting Batteries 14 7 Service Checkout Pro...

Страница 4: ...ructions for Use Serial Number Manufacturer Abbreviations Used In This Manual CCW Counter clockwise decrease CW Clockwise increase CVR Continuous Vacuum Regulator MAX Full Line Vacuum inHg Inches of m...

Страница 5: ...ter will help prevent this and extend the life of suction equipment 1 3 CAUTION Do not use any Loctite products or any products which contain Methacrylate Ester as an active ingredient to seal the thr...

Страница 6: ...00 kPa mm Hg Analog 1230 Digital 1337 High Two Mode Continuous Adult L L M M H H F F MEDICAL VACUUM Analog 1330 Digital 1430 High Two Mode Continuous Adult mm Hg 100 200 500 700 300 400 600 MAX L L M...

Страница 7: ...uum is provided In the regulated mode the vacuum source is connected through the regulator module which functions as an automatic valve Turning the suction control knob adjusts the position of the reg...

Страница 8: ...l scale 1 Full scale High 0 760 mmHg 0 101 3 KPA 5 Full scale 1 Full scale Pediatric 0 160 mmHg 0 21 3 KPA 5 Full scale 1 Full scale Neonatal 0 160 mmHg 0 21 3 KPA 0 100 mmHg 0 13 3 KPA 5 Full scale 1...

Страница 9: ...ine system Failure to clean and disinfect the wall outlet and pipeline system may result in damage to this equipment 4 1 1 Routine Exterior Cleaning Routine cleaning of the regulator is recommended as...

Страница 10: ...ated products DO NOT SEND UNIT BACK TO THE MANUFACTURER Gauge doesn t respond to changes in suction Gauge assembly is not properly aligned Ensure gauge assembly 6 is properly aligned Missing Damaged O...

Страница 11: ...e assembly Replace gauge assembly 6 Whistling buzzing noise from the unit Need to apply grease on stem on regulator module Apply grease on stem 31 Gauge responds slowly to changes in suction failed bl...

Страница 12: ...est the gauge on its face Note See Section 9 1 Illustrated Parts 1 Remove the four cover screws from the back of the regulator 2 With regulator facing up carefully pull the cover assembly off the back...

Страница 13: ...with a drop of removable thread locker such as Loctite 242 Repeat step 5 to verify the vacuum relief valve setting 8 Re attach the cover see Section 6 2 2 Assembly WARNING Excess Loctite may seal the...

Страница 14: ...Place the O ring on the housing 5 Fold the diaphragm and place it on the piston head 6 Place the cap on the housing and press until it snaps into place 6 5 Digital Gauge Note Both batteries are positi...

Страница 15: ...ith an accuracy of 1 of full range at 22 C Note All gauge needles should come to rest within the zero range bracket or return to the stop pin when no suction is being supplied Gauges which do not comp...

Страница 16: ...1 mmHg 0 1 kPa for the NEO model and 8 mmHg 1 0 kPa for the high model 7 5 Maximum Vacuum Test PED NEO C A S S Models ONLY 1 Occlude the fitting patient port 2 Rotate the suction control knob fully c...

Страница 17: ...ting patient and adapter probe ports with minimum inner diameters of 0 140 The unit is either directly connected to the vacuum source or connected with a tube no more than 4 in length with a minimum i...

Страница 18: ...nvironmental conditions Replace lubricate and repair as needed 8 3 Repair Policy WARNING To reduce transportation personnel and or service personnel exposure to hazardous contamination clean and disin...

Страница 19: ...nd Fittings CAUTION Do not use any Loctite products or any products which contain Methacrylate Ester as an active ingredient to seal the threads on the adapters probes and fittings All Ohio Medical re...

Страница 20: ...acuum Relief Assembly 1 36 8700 0024 500 Quad Ring 006 Nitrile 1 18 See table below O Ring 2 014 Nitrile Buna N 1 8700 0006 000 PTS CVR Service Manual 1 19 6700 0115 400 Relief Valve Positive Pressure...

Страница 21: ...21 8700 0006 000 Rev 12 08 2020 9 Ordering Information 23 15 14 13 11 1 5 4X 7 4X 6 5 4 7 7 25 2 26 3 27 1 12 16 10 8 5 18 19 20 21 22 17 24 32 31 30 29 28 35 36 34 33...

Страница 22: ...hobic Filter 1 8 NPT x Tubing nipple 6730 0580 800 3 Pack 6730 0581 800 10 Pack 6730 0582 800 50 Pack Tubing 6700 0005 300 Reusable Overflow Safety Traps Description Part Number Overflow Safety Trap w...

Страница 23: ...C 61000 4 2 8kV Contact 2 kV 4 kV 8 kV and 15 kV Air 8kV Contact 2 kV 4 kV 8 kV and 15 kV Air Floors should be wood concrete or ceramic tile If the floors are covered with synthetic material the relat...

Страница 24: ...unity Test Level V m 385 380 390 TETRA 400 Pulse modulation 18 Hz 1 8 0 3 27 450 430 470 GMRS 460 FRS460 FM 5 kHz deviation 1 kHz sine 2 0 3 28 710 704 787 LTE Band 13 17 Pulse modulation 217 Hz 0 2 0...

Страница 25: ...x is a registered trademark of S C Johnson Son Inc Kleenaseptic is a registered trademark of Metrex Research Corporation Betco is a registered trademark of Betco Corporation Cavicide is a registered t...

Отзывы: