CHAPTER 1  GETTING STARTED 

Valor

 7000 Series Service Manual

  

1-1

 

Ohaus Corporation  www.ohaus.com

 

1.1  INTRODUCTION 

This service manual contains the information needed to perform routine maintenance and 
service on the Ohaus Valor 7000 Series scales.  

Familiarity  with  the  scale’s  Instruction  

Manual is assumed.  The contents of this manual are contained in five chapters: 

Chapter 1  Getting Started

 

 Contains information on service facilities, tools and test 

equipment, specifications, and the mechanical and electronic functions of the scale. 

Chapter 2  Troubleshooting

 

 Contains a diagnostic guide and error code table. 

Chapter 3  Maintenance Procedures

 

 Contains preventive maintenance procedures and 

disassembly, repair and replacement procedures. 

Chapter 4  Testing

 

 Contains a list of required test masses, an operational test, segment 

display test, performance tests and adjustments. 

Chapter 5  Drawings and Parts

 Lists 

 Contains exploded views of Valor 7000 scales 

identifying all serviceable components. 

Appendix A  Standard Calibration

 

 Explains procedures for Standard Calibration, 

performed prior to using a scale, and after service. 

Appendix B  Service Calibration

 

 Describes the Service Menu and sub-menus, which 

allow authorized service personnel to perform factory Linearity and Span calibrations (no 
pre-set limits), take Ramp readings, and to set the service menu units. 

1.2  SERVICE FACILITIES 

To service a scale, the service area should meet the following requirements: 

x

  Should be temperature controlled and meet scale specifications for temperature 

environmental requirements. 

x

  Must be free of vibrations such as fork lift trucks close by, large motors, air currents 

or drafts from air conditioning/heating ducts, open windows, people walking by, fans, 
etc. 

x

  Area must be clean and free of excessive dust.  

x

  Work surface must be stable and level. 

x

  Scale must not be exposed to direct sunlight or radiating heat sources. 

x

  Use an approved Electro-Static Device. 

 

1.3  TOOLS AND TEST EQUIPMENT REQUIRED 

 

The service shop should contain the following equipment: 

1.  Standard hand tools. 
2.  Digital Voltmeter (DVM). 
3.  Standard Electronics tool kit. 
4.  Grounding mat and clip. 
5.  Strain Gauge Simulator. 

 

 

 

Содержание Valor 7000 series

Страница 1: ...SERVICE MANUAL VALOR 7000 Ohaus Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 973 377 9000...

Страница 2: ...publication but Ohaus Corporation assumes no liability arising from the use or misuse of this material Reproduction of this material is strictly prohibited Material in this manual is subject to chang...

Страница 3: ...3 Testing the Printed Circuit Board PCB 2 4 2 2 4 Diagnostic Guide 2 6 CHAPTER 3 MAINTENANCE PROCEDURES 3 1 Preventive Maintenance 3 1 3 1 1 Preventive Maintenance Checklist 3 1 3 2 Service Strategy...

Страница 4: ...ucture 1 5 2 1 Load Cell Resistance Readings in Ohms 2 2 2 2 Color Code for Load Cell Wiring 2 2 2 3 Load Cell Output Readings in mV with 3 3V Excitation 2 3 2 4 Diagnostic Guide 2 6 3 1 Mounting Bolt...

Страница 5: ...ponents Appendix A Standard Calibration Explains procedures for Standard Calibration performed prior to using a scale and after service Appendix B Service Calibration Describes the Service Menu and su...

Страница 6: ...0 1 30000 LFT ON Resolution nMax 1 3000 1 3000 1 3000 1 3000 1 3000 Linearity 2d r0 0001 kg r0 0002 kg r0 0005 kg r0 001 kg r0 002 kg Repeatability 2d r0 0001 kg r0 0002 kg r0 0005 kg r0 001 kg r0 002...

Страница 7: ...g Press Toggles through active weighing units Short Press in Menu Toggles through available settings Function Mode Back Short Press Selects function setting Long Press Selects active Mode Short Press...

Страница 8: ...item 11 in figure 3 2 will remain lit The scale can be operated during charging and the battery is protected against overcharging For maximum operating time the battery should be charged at room tempe...

Страница 9: ...O M L O C k E n d Menu Items Reset kg g lb oz lb oz End Reset Stable A Print Contnt Layout Data Tr End Reset Baud Parity Stop Handsh Alt Cm End L Cal L Setup L Read L Mode L Unit L Print L COM End 1 5...

Страница 10: ...is not accessible Zero Range is locked at 2 Stable Range setting is locked at 1d Auto Zero Tracking setting is locked at 0 5d Units are locked at their current settings Filter is locked at current se...

Страница 11: ...CHAPTER 1 GETTING STARTED Valor 7000 Series Service Manual 1 7 Ohaus Corporation www ohaus com Figure 1 3 Sealing...

Страница 12: ...ith the most common cause first If the first remedy does not fix the problem proceed to the next remedy Before attempting to repair the balance read all chapters of this manual to be familiar with the...

Страница 13: ...d full load output The reading should meet the Load Cell specifications The Load Cell output should be very close to linear over its capacity range NOTE The following steps involve power applied to th...

Страница 14: ...ts at Zero Load 50 and full scale capacities See Table 2 3 for typical readings Figure 2 1 Top view of main PCB NOTE Table 2 3 indicates typical readings Actual values can vary but should remain linea...

Страница 15: ...n 3 Using a DVM measure the excitation voltage EXE and EXE in Figure 2 1 The reading should be 3 1 volts dc This is the excitation voltage for the Load Cell and is regulated If the voltage is lower re...

Страница 16: ...ue 6 Adjust the Simulator to simulate 0 load 50 load and 100 load for the capacity that the scale is rated for See Table 2 3 for values to use If the resulting readings are unstable the Main PC Board...

Страница 17: ...below Power On Zero limit Err 8 3 Over Range Error Weight reading exceeds Overload limit Err 8 4 Under Range Error Weight reading below Underload limit Err 8 5 Tare out of range Adjust tare value to b...

Страница 18: ...tronics to stabilize 3 1 1 Preventive Maintenance Checklist The scale should be inspected and checked regularly as follows 1 Remove the Pan and Sub Pan to inspect and clean the area beneath the Pan 2...

Страница 19: ...the eight screws holding the Housing in place See Figure 3 1 3 Move the top housing a little bit backward and then raise and remove the top housing 4 Separate Top Housing from Bottom Housing Avoid str...

Страница 20: ...PCB to the bottom housing 3 If either the Main PCB or the Load Cell is to be replaced note the order of the wire colors and then disconnect the Load Cell Cable from the Main PCB See Figure 3 2 4 Posit...

Страница 21: ...washers that hold the Load Cell assembly frame and load cell to the Bottom Housing Remove the bolts and their washer pairs that hold the Load Cell Frame to the Load Cell Use a high leverage Allen wren...

Страница 22: ...ual to the scale s capacity See Table 3 2 There are four Overload Stop Bolts See Figure 3 5 Adjust the Overload Stops per Table 3 2 1 Adjust the Overload Stop Bolt so that the gap between the Load Cel...

Страница 23: ...VERLOAD STOP GAP SETTINGS Model Max Capacity Overload Stop Gap V7000 Scale Load Cell A B mm C D mm V71P1501T 1 5kg 3kg 0 6 0 5 V71P3T 3kg 5kg 0 8 0 7 V71P6T 6kg 10kg 1 5 1 1 V71P15T 15kg 20kg 2 4 1 7...

Страница 24: ...2 Pull the quick connect tabs free of the battery posts and lift out the battery Figure 3 7 Rechargeable battery 3 Position the replacement battery in the compartment on an angle sufficient to access...

Страница 25: ...S REQUIRED The masses required to test the Ohaus Valor 7000 scales must meet the requirements of ASTM Class 4 or OIML F2 Tolerance The mass values are listed in Table 4 1 TABLE 4 1 TEST MASS VALUES Mo...

Страница 26: ...ication for each model listed in Table 1 1 1 Power on the balance The reading on the display should be 0g 2 Select a mass weighing near the maximum capacity of the balance and place it on the center o...

Страница 27: ...test set up the instrument as follows Set Up Follow the steps in Appendix A Section A 2 Setup and Calibration Record Settings Zero Tracking Setting ____________ Displayed Units ____________ Mass Used...

Страница 28: ...ge 6 Mean Reading 1 Reading 2 Reading 3 Reading 4 Reading 5 7 Reading 6 Reading 7 Reading 8 Reading 9 Reading 10 10 Mean ________ 6 Calculate the Delta for each reading and record in the work sheet De...

Страница 29: ...press ON Zero Off 3 Place the test mass on the Scale record the weight and remove 4 Place Load 2 on the Scale and press ON Zero Off 5 Place the test mass on the Scale record the weight and remove 6 P...

Страница 30: ...test weight in the center of the Weighing Pan 2 Tare the balance 3 Move the weight to point A and record the reading 4 Move the weight to point B and record the reading 5 Move the weight to point C an...

Страница 31: ...d NOTE In all cases where a part is replaced the scale must be thoroughly checked after the replacement is made The scale MUST meet the parameters of all applicable specifications in this manual If fu...

Страница 32: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 2 Valor 7000 Series Service Manual 5 1 Valor 7000 SCALES PARTS Figure 5 1 Valor 7000 Scales Parts...

Страница 33: ...der Down R31 R21 RC31 RC21 V71 8 30037420 Hardware Kit R31 R21 RC31 RC21 V71 9 30037400 Housing Bottom V71 10 30037421 Feet R31 R21 RC31 RC21 V71 11 30037422 Cover Option R31 R21 RC31 RC21 V71 12 3003...

Страница 34: ...libration process 2 Allow the scale to warm up for approximately five minutes after stabilizing to room temperature 3 To abort calibration press Exit key or turn the scale off anytime during the calib...

Страница 35: ...accept SpaN will then be shown Press Yes to begin the span calibration 0 kg will be displayed SpaN 0 3 Press Yes to accept C will be displayed while zero reading is stored Next the display shows the c...

Страница 36: ...factory Linearity and Span calibrations no pre set limits Note Be careful not to touch the scale or the surface it rests on while calibration is in progress as it will cause the process to fail B 1 S...

Страница 37: ...e Service Span Calibration menu Set the graduation of the scale 0 00005 0 02 Grad B 5 Service Span Calibration When SPAN appears press YES to accept the Service Span Calibration menu or No to advance...

Страница 38: ...to indicate the calibration unit The example shows the mid calibration point value for a Scale Capacity of 3 kg Place the specified calibration weight on the pan and press Yes C flashes while the mid...

Страница 39: ...o the Expand menu item Note Ramp is used to troubleshoot load cell problems 0 represents no signal from the A D 100 reflects the upper signal range of the A D ramMp 0 0 to 100 0 B 8 Expand Menu When e...

Отзывы: