background image

CHAPTER 4  TESTING 

Valor

 7000 Series Service Manual 

4-3 

Ohaus Corporation  www.ohaus.com 

4.4.2  Repeatability Test 

Repeatability is the Standard Deviation of a set of similar weight readings. 

Requirements: 

  To perform this test a single mass must be used for all readings. 

  The test mass should be approximately 

½

 of the capacity of the instrument. 

  Wear gloves when handling the mass. 

Before starting a repeatability test, set up the instrument as follows: 

Set Up: 

Follow the steps in Appendix A, Section A-2, Setup and Calibration. 

Record Settings: 

Zero Tracking Setting =   ____________ 

Displayed Units =   

____________ 

Mass Used =  

 

____________

 

TEST PROCEDURE: 

1.  Zero the instrument, if it does not read zero. 

2.  Using a test mass approximately half the capacity of the instrument, place the mass on 

the center of platform. Record the reading on the worksheet provided. 

3.  Remove the mass from the platform. 

4.  Repeat this test starting at Step 1 until you record a total of ten readings 

Fill in the worksheet (Table 4-2) with the ten (10) readings. 

Содержание Valor 7000 series

Страница 1: ...SERVICE MANUAL VALOR 7000 Ohaus Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 973 377 9000...

Страница 2: ...publication but Ohaus Corporation assumes no liability arising from the use or misuse of this material Reproduction of this material is strictly prohibited Material in this manual is subject to chang...

Страница 3: ...3 Testing the Printed Circuit Board PCB 2 4 2 2 4 Diagnostic Guide 2 6 CHAPTER 3 MAINTENANCE PROCEDURES 3 1 Preventive Maintenance 3 1 3 1 1 Preventive Maintenance Checklist 3 1 3 2 Service Strategy...

Страница 4: ...ucture 1 5 2 1 Load Cell Resistance Readings in Ohms 2 2 2 2 Color Code for Load Cell Wiring 2 2 2 3 Load Cell Output Readings in mV with 3 3V Excitation 2 3 2 4 Diagnostic Guide 2 6 3 1 Mounting Bolt...

Страница 5: ...ponents Appendix A Standard Calibration Explains procedures for Standard Calibration performed prior to using a scale and after service Appendix B Service Calibration Describes the Service Menu and su...

Страница 6: ...0 1 30000 LFT ON Resolution nMax 1 3000 1 3000 1 3000 1 3000 1 3000 Linearity 2d r0 0001 kg r0 0002 kg r0 0005 kg r0 001 kg r0 002 kg Repeatability 2d r0 0001 kg r0 0002 kg r0 0005 kg r0 001 kg r0 002...

Страница 7: ...g Press Toggles through active weighing units Short Press in Menu Toggles through available settings Function Mode Back Short Press Selects function setting Long Press Selects active Mode Short Press...

Страница 8: ...item 11 in figure 3 2 will remain lit The scale can be operated during charging and the battery is protected against overcharging For maximum operating time the battery should be charged at room tempe...

Страница 9: ...O M L O C k E n d Menu Items Reset kg g lb oz lb oz End Reset Stable A Print Contnt Layout Data Tr End Reset Baud Parity Stop Handsh Alt Cm End L Cal L Setup L Read L Mode L Unit L Print L COM End 1 5...

Страница 10: ...is not accessible Zero Range is locked at 2 Stable Range setting is locked at 1d Auto Zero Tracking setting is locked at 0 5d Units are locked at their current settings Filter is locked at current se...

Страница 11: ...CHAPTER 1 GETTING STARTED Valor 7000 Series Service Manual 1 7 Ohaus Corporation www ohaus com Figure 1 3 Sealing...

Страница 12: ...ith the most common cause first If the first remedy does not fix the problem proceed to the next remedy Before attempting to repair the balance read all chapters of this manual to be familiar with the...

Страница 13: ...d full load output The reading should meet the Load Cell specifications The Load Cell output should be very close to linear over its capacity range NOTE The following steps involve power applied to th...

Страница 14: ...ts at Zero Load 50 and full scale capacities See Table 2 3 for typical readings Figure 2 1 Top view of main PCB NOTE Table 2 3 indicates typical readings Actual values can vary but should remain linea...

Страница 15: ...n 3 Using a DVM measure the excitation voltage EXE and EXE in Figure 2 1 The reading should be 3 1 volts dc This is the excitation voltage for the Load Cell and is regulated If the voltage is lower re...

Страница 16: ...ue 6 Adjust the Simulator to simulate 0 load 50 load and 100 load for the capacity that the scale is rated for See Table 2 3 for values to use If the resulting readings are unstable the Main PC Board...

Страница 17: ...below Power On Zero limit Err 8 3 Over Range Error Weight reading exceeds Overload limit Err 8 4 Under Range Error Weight reading below Underload limit Err 8 5 Tare out of range Adjust tare value to b...

Страница 18: ...tronics to stabilize 3 1 1 Preventive Maintenance Checklist The scale should be inspected and checked regularly as follows 1 Remove the Pan and Sub Pan to inspect and clean the area beneath the Pan 2...

Страница 19: ...the eight screws holding the Housing in place See Figure 3 1 3 Move the top housing a little bit backward and then raise and remove the top housing 4 Separate Top Housing from Bottom Housing Avoid str...

Страница 20: ...PCB to the bottom housing 3 If either the Main PCB or the Load Cell is to be replaced note the order of the wire colors and then disconnect the Load Cell Cable from the Main PCB See Figure 3 2 4 Posit...

Страница 21: ...washers that hold the Load Cell assembly frame and load cell to the Bottom Housing Remove the bolts and their washer pairs that hold the Load Cell Frame to the Load Cell Use a high leverage Allen wren...

Страница 22: ...ual to the scale s capacity See Table 3 2 There are four Overload Stop Bolts See Figure 3 5 Adjust the Overload Stops per Table 3 2 1 Adjust the Overload Stop Bolt so that the gap between the Load Cel...

Страница 23: ...VERLOAD STOP GAP SETTINGS Model Max Capacity Overload Stop Gap V7000 Scale Load Cell A B mm C D mm V71P1501T 1 5kg 3kg 0 6 0 5 V71P3T 3kg 5kg 0 8 0 7 V71P6T 6kg 10kg 1 5 1 1 V71P15T 15kg 20kg 2 4 1 7...

Страница 24: ...2 Pull the quick connect tabs free of the battery posts and lift out the battery Figure 3 7 Rechargeable battery 3 Position the replacement battery in the compartment on an angle sufficient to access...

Страница 25: ...S REQUIRED The masses required to test the Ohaus Valor 7000 scales must meet the requirements of ASTM Class 4 or OIML F2 Tolerance The mass values are listed in Table 4 1 TABLE 4 1 TEST MASS VALUES Mo...

Страница 26: ...ication for each model listed in Table 1 1 1 Power on the balance The reading on the display should be 0g 2 Select a mass weighing near the maximum capacity of the balance and place it on the center o...

Страница 27: ...test set up the instrument as follows Set Up Follow the steps in Appendix A Section A 2 Setup and Calibration Record Settings Zero Tracking Setting ____________ Displayed Units ____________ Mass Used...

Страница 28: ...ge 6 Mean Reading 1 Reading 2 Reading 3 Reading 4 Reading 5 7 Reading 6 Reading 7 Reading 8 Reading 9 Reading 10 10 Mean ________ 6 Calculate the Delta for each reading and record in the work sheet De...

Страница 29: ...press ON Zero Off 3 Place the test mass on the Scale record the weight and remove 4 Place Load 2 on the Scale and press ON Zero Off 5 Place the test mass on the Scale record the weight and remove 6 P...

Страница 30: ...test weight in the center of the Weighing Pan 2 Tare the balance 3 Move the weight to point A and record the reading 4 Move the weight to point B and record the reading 5 Move the weight to point C an...

Страница 31: ...d NOTE In all cases where a part is replaced the scale must be thoroughly checked after the replacement is made The scale MUST meet the parameters of all applicable specifications in this manual If fu...

Страница 32: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 2 Valor 7000 Series Service Manual 5 1 Valor 7000 SCALES PARTS Figure 5 1 Valor 7000 Scales Parts...

Страница 33: ...der Down R31 R21 RC31 RC21 V71 8 30037420 Hardware Kit R31 R21 RC31 RC21 V71 9 30037400 Housing Bottom V71 10 30037421 Feet R31 R21 RC31 RC21 V71 11 30037422 Cover Option R31 R21 RC31 RC21 V71 12 3003...

Страница 34: ...libration process 2 Allow the scale to warm up for approximately five minutes after stabilizing to room temperature 3 To abort calibration press Exit key or turn the scale off anytime during the calib...

Страница 35: ...accept SpaN will then be shown Press Yes to begin the span calibration 0 kg will be displayed SpaN 0 3 Press Yes to accept C will be displayed while zero reading is stored Next the display shows the c...

Страница 36: ...factory Linearity and Span calibrations no pre set limits Note Be careful not to touch the scale or the surface it rests on while calibration is in progress as it will cause the process to fail B 1 S...

Страница 37: ...e Service Span Calibration menu Set the graduation of the scale 0 00005 0 02 Grad B 5 Service Span Calibration When SPAN appears press YES to accept the Service Span Calibration menu or No to advance...

Страница 38: ...to indicate the calibration unit The example shows the mid calibration point value for a Scale Capacity of 3 kg Place the specified calibration weight on the pan and press Yes C flashes while the mid...

Страница 39: ...o the Expand menu item Note Ramp is used to troubleshoot load cell problems 0 represents no signal from the A D 100 reflects the upper signal range of the A D ramMp 0 0 to 100 0 B 8 Expand Menu When e...

Отзывы: