APPENDIX A STANDARD CALIBRATION
APPENDIX A.
STANDARD CALIBRATION
A.1 CALIBRATION
Note
: This menu is locked out in Legal For Trade applications. To regain access,
see Section 1.6.
Standard calibration should be performed prior to using a scale, and after service.
See Section 4.2 for Calibration Masses required for each model.
Note:
Be careful not to touch the scale or the table while calibration is in progress,
as it will cause the process to fail.
1. Press and hold the
ENTER/Ca
l
button
until
.C.A.L.
appears.
.C.a.l.
2. Press
Yes
.
CAL
appears, followed by
a flashing
-C-
while a zero reading is
stored.
-C-
3. The specified calibration weight value
appears. Place the weight on the pan.
The display shows
-C-
while the
reading is stored.
6000
4. When calibration is complete, the
display shows done, and then returns
to the previous application mode and
scale is ready for use.
Done
NOTE:
If calibration fails, ensure that the test area is free from drafts and the surface the scale
rests on is level and free of vibrations. Then try to calibrate again. If it continues to fail, there
may be an internal problem. To resolve internal problems, follow procedures in Chapter 3.
Valor™ 3000 Series Service Manual
A-1
Ohaus Corporation www.ohaus.com
Содержание Valor 3000
Страница 2: ......
Страница 4: ......
Страница 8: ...TABLE OF CONTENTS Valor 3000 Series Service Manual iv Ohaus Corporation www ohaus com...
Страница 16: ...CHAPTER 1 GETTING STARTED Ohaus Corporation www ohaus com 1 8 Valor 3000 Series Service Manual...
Страница 36: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 6 Valor 3000 Series Service Manual...
Страница 38: ...APPENDIX A STANDARD CALIBRATION Ohaus Corporation www ohaus com A 2 Valor 3000 Series Service Manual...
Страница 44: ...APPENDIX B THE SERVICE MENU Ohaus Corporation www ohaus com B 6 Valor 3000 Series Service Manual...
Страница 51: ......
Страница 52: ...80252591 P N 80152591 SERVICE MANUAL Valor 3000 SERIES SCALES...