CHAPTER 4 TESTING
4.5
Performance Tests
Accurate performance of the Valor 3000 scales is determined by a series of four performance
tests. The displayed readings are compared with the tolerances listed in Tables 4-2 and 4-3.
Tolerance values are expressed in counts. A one-count difference is shown in the last digit on
the scale display.
TABLE 4-2. TOLERANCES
Model
V31XH202 V31XH2 V31X3 V31X6 V31XW301
V31XW3/
V31X3N
V31XW6/
V31X3N
Precision
±1 ±1 ±1 ±1 ±2 ±1 ±2
Repeatability
1
1
1
1
2
1
2
Linearity
±1 ±1 ±1 ±1 ±2 ±1 ±2
NOTE:
The following performance tests are used to evaluate scale operation before and
after repairs. The scale must meet the requirements specified in each test as
well as the specifications listed in Table 4-1. Before proceeding with the
following tests, the scale should be calibrated. (See Appendix A.)
4.5.1 Precision Test
The Precision Test measures the Standard Deviation of a set of similar weight readings, which
should match the specification for each model, listed in Tolerance Table 4-2.
1. Power on the balance. The reading on the display should be 0g.
2. Select a mass weighing near the maximum capacity of the balance, and place it on
the center of the Pan. Observe and record the reading.
3. Remove the mass. The reading should return to 0g ± the tolerance in Table 4-2.
4. Repeat this test three times. The readings should be within tolerances. If so, the
balance passes the Precision Test.
5. If the deviation for any set of readings (using the same mass placed on the center of
the Pan) is greater than the tolerance listed in Table 4-2, the balance does not meet
the precision specification. Inspect and correct the following areas:
– Check for mechanical obstructions. Any foreign object touching any part of the
moving assemblies will cause a balance to fail the Precision Test. Inspect and
correct as necessary.
– If the scale does not meet specifications, move it to a suitable location, ensure
that it is level, and try again. If it still does not meet specifications, perform a
service calibration, and try again. (See Appendix B for Service Calibration.)
– If the scale does not pass this test, the Load Cell may need to be replaced.
Ohaus Corporation www.ohaus.com
4-2
Valor™ 3000 Series Service Manual
Содержание Valor 3000
Страница 2: ......
Страница 4: ......
Страница 8: ...TABLE OF CONTENTS Valor 3000 Series Service Manual iv Ohaus Corporation www ohaus com...
Страница 16: ...CHAPTER 1 GETTING STARTED Ohaus Corporation www ohaus com 1 8 Valor 3000 Series Service Manual...
Страница 36: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 6 Valor 3000 Series Service Manual...
Страница 38: ...APPENDIX A STANDARD CALIBRATION Ohaus Corporation www ohaus com A 2 Valor 3000 Series Service Manual...
Страница 44: ...APPENDIX B THE SERVICE MENU Ohaus Corporation www ohaus com B 6 Valor 3000 Series Service Manual...
Страница 51: ......
Страница 52: ...80252591 P N 80152591 SERVICE MANUAL Valor 3000 SERIES SCALES...