CHAPTER 1 GETTING STARTED
1.6 LEGAL FOR TRADE (LFT)
When the scale is set for LFT (on XW and NTEP models), the following menu settings cannot
be accessed/changed: Calibration, Unit, Stability, AST, Filter, Power On Unit, and Global Reset.
To regain access to the locked menu settings, remove the small rectangular LFT/Calibration
Lock Switch cover on the bottom of the scale. (See Figure 1-4.) Use a small flathead screw
driver to move the CAL switch to the left (closer to the “L” in “CAL”), which turns LFT off.
After the LFT/CAL Lock setting has been
turned off, the scale must be inspected in
accordance with local weights and measures
regulations before it can be used in LFT mode
again. Local authorities may apply seals on
the three screws indicated in Figure 1-4.
Figure 1-4. LFT/Cal Lock Switch.
Screws to
which LFT
seals may
be applied
1.6.1 LFT SETUP PROCEDURE
4. Set the Geographical Adjustment Factor (GEO) according to the location where the
scale is to be used. (See Appendix D. Possible values are from 1 to 31.) The
setting is in the Service Menu: press
O/T/Off
and
MODE/UNITMenu
simultaneously
during power on. Press
No
three times until GEO
appears, then press
Yes
. Then
continuously press
No
until the desired number appears. Press
Yes
, then
No
until
END appears. Press
Yes
to return to weighing mode.
5. Access the Setup menu (Press and hold
MODE/UNIT/Menu
until SEtUP appears.
Press
Yes
to enter the displayed menu, or
No
to advance to the next menu, etc.)
Set the following:
Auto Zero Tracking (AZT) = 0.5d
Stability (Stab) =1d
Units as per local LFT requirements
6. Perform an accurate calibration. (See Appendix A.)
7. Turn the Scale off
and disconnect
the AC power connection
.
8. Open the rectangular LFT/Calibration Lock Switch Cover on the bottom of the scale
.
9. Move the switch to the right (closer to the “C” in “CAL”), which turns LFT on.
10. Reinstall the LFT/Calibration Lock Switch Cover.
11. Reconnect the AC power, press
O/T/On/Off
to turn on the scale. During power on,
the display sequence will now include: “LFT.” This message confirms the proper
setting of the LFT mode and the hardware lock switch.
Ohaus Corporation www.ohaus.com
1-6
Valor™ 3000 Series Service Manual
Содержание Valor 3000
Страница 2: ......
Страница 4: ......
Страница 8: ...TABLE OF CONTENTS Valor 3000 Series Service Manual iv Ohaus Corporation www ohaus com...
Страница 16: ...CHAPTER 1 GETTING STARTED Ohaus Corporation www ohaus com 1 8 Valor 3000 Series Service Manual...
Страница 36: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 6 Valor 3000 Series Service Manual...
Страница 38: ...APPENDIX A STANDARD CALIBRATION Ohaus Corporation www ohaus com A 2 Valor 3000 Series Service Manual...
Страница 44: ...APPENDIX B THE SERVICE MENU Ohaus Corporation www ohaus com B 6 Valor 3000 Series Service Manual...
Страница 51: ......
Страница 52: ...80252591 P N 80152591 SERVICE MANUAL Valor 3000 SERIES SCALES...