CHAPTER 3 REPAIR PROCEDURES
Ohaus Corporation www.ohaus.com
3-8 3000 Series T32ME & MC Indicators Service Manual
3.6 FUNCTION
LABEL
REPLACEMENT
1. Carefully remove the old Function Label. It is held in place with a strong adhesive.
2. Clean the Top Cover and remove all traces of adhesive.
3. Remove the protective backing from the back of the new Function Label.
Carefully position it on the Top Housing, starting at the bottom of the cover.
Use a rolling motion to smooth it into position.
4. Close the Housing (if opened previously), re-insert and tighten the four screws
previously removed.
5. Perform calibration procedures. (See Appendix A.)
6.
Perform testing procedures. (See Chapter 4.)
Содержание T32ME
Страница 2: ......
Страница 4: ......
Страница 8: ...TABLE OF CONTENTS Ohaus Corporation www ohaus com iv 3000 Series T32ME MC Indicators Service Manual...
Страница 22: ...CHAPTER 2 TROUBLESHOOTING Ohaus Corporation www ohaus com 2 4 3000 Series T32ME MC Indicators Service Manual...
Страница 45: ......
Страница 46: ...80253153 P N 80253153 SERVICE MANUAL 3000 SERIES T32MC T32ME INDICATORS...