CHAPTER 5  PARTS LISTS & DIAGRAMS 

Ohaus Corporation  www.ohaus.com 

5-6  3000 Series T32ME & MC Indicators Service Manual 

 

Содержание T32ME

Страница 1: ...SERVICE MANUAL 3000 Series T32ME T32MC Indicators T32ME T32MC Ohaus Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 973 377 9000...

Страница 2: ......

Страница 3: ...ate at the time of publication but Ohaus Corporation assumes no liability arising from the use or misuse of this material Reproduction of this material is strictly prohibited Material in this manual i...

Страница 4: ......

Страница 5: ...G 2 1 Introduction 2 1 2 2 Visual Inspection 2 1 2 3 Error Codes 2 2 2 4 Problem Solver 2 3 CHAPTER 3 MAINTENANCE PROCEDURES 3 1 Preventive Maintenance 3 1 3 1 1 Preventive Maintenance Checklist 3 1 3...

Страница 6: ...Parts 5 3 5 2 T32ME LED Indicator Housing Internal Parts 5 5 A 1 Setup Calibration Values A 3 B 1 Geographical Adjustment Values B 1 LIST OF ILLUSTRATIONS FIGURE NO TITLE 1 1 Four screws on rear panel...

Страница 7: ...T32ME MC Indicators Service Manual iii Ohaus Corporation www ohaus com FIGURE NO TITLE Page No 3 11 Main PCB 3 6 5 1 Series 3000 T32MC Indicator Housing Internal Parts 5 2 5 1 Series 3000 T32ME Indica...

Страница 8: ...TABLE OF CONTENTS Ohaus Corporation www ohaus com iv 3000 Series T32ME MC Indicators Service Manual...

Страница 9: ...ational test segment display test performance tests and adjustments Chapter 5 Drawings and Parts Lists Contains exploded views of T32 Indicators identifying all serviceable components Before servicing...

Страница 10: ...pecial Tools and Test Equipment List 1 Ohaus Scale Base 2 Load Cell Simulator optional 3 Computer with RS232 Interface for testing the RS232 communications 4 RS232 Interface cable 25 pin PN 80500524 9...

Страница 11: ...digit 7 segment backlit LCD 0 8 20mm digit height 6 digit 7 segment Red LED Backlight White LED Keypad 4 button membrane switches Load Cell Excitation Voltage 5V DC Load Cell Drive Up to 4 x 350 ohm l...

Страница 12: ...the threaded knobs holding the Indicator by hand CAUTION ELECTRICAL SHOCK HAZARD REMOVE ALL POWER CONNECTIONS TO THE INDICATOR BEFORE SERVICING OR MAKING INTERNAL CONNECTIONS Remove the four screws fr...

Страница 13: ...Jumpers W1 and W2 For load cells with a ground shield wire connect the shield to the center position GND Figure 1 4 Jumpers 1 5 2 RS232 Interface Each model has a standard 9 pin D style RS232 connect...

Страница 14: ...ges the weighing unit Mode Allows changing the application mode Press and hold allows scrolling through modes Menu Enter the User menu Extended press View the Audit Trail event counters Menu Function...

Страница 15: ...Menu Mode press and hold the TARE Menu button until MENU appears on the display The first upper level menu appears on the display Use Yes No Back and Exit to navigate in Menu Mode Yes Allows entry int...

Страница 16: ...No first 7 Even then 7 Odd 7 None and 8 None again Press Yes to accept a setting PAritY appears again Press No 6 StOP appears Press Yes The default Stop Bit setting 1 appears To change the setting pre...

Страница 17: ...t g kg lb oz lb oz xU Set scale to unit x 1 g 2 kg 3 lb 4 oz 5 lb oz PV Version print name software revision and LFT ON if LFT is set ON Esc R Global reset to reset all menu settings to the original f...

Страница 18: ...set the LFT menu item to OFF as follows 1 Press and hold TARE Menu until C A L appears then press No 2 S E t U P appears Press Yes 3 rESEt appears Press No 4 LFt appears Press Yes 5 OFF appears Press...

Страница 19: ...note their condition 1 Examine the Indicator for signs of abuse such as a cracked top cover or base damage from liquids cracks on corners which may suggest the unit was dropped 2 Open the Indicator by...

Страница 20: ...necessary Error 8 3 Weight reading exceeds Overload limit Reduce load on scale If error continues Test the PCB replace if necessary Error 8 4 Weight reading below Underload limit Recalibrate If error...

Страница 21: ...le exceeds allowable limits Load on Scale is not stable Load Cell is damaged PCB is damaged Remove load on Scale Wait for load to become stable Replace Load Cell Test Replace PCB Section 3 6 Poor accu...

Страница 22: ...CHAPTER 2 TROUBLESHOOTING Ohaus Corporation www ohaus com 2 4 3000 Series T32ME MC Indicators Service Manual...

Страница 23: ...side using a cloth dampened with a mild detergent CAUTION DO NOT USE CHEMICAL CLEANERS OR SOLVENTS OF ANY TYPE SOME CLEANERS ARE ABRASIVE AND MAY AFFECT THE SCALE S FINISH 3 Check the Power Cord for b...

Страница 24: ...wer supply jack Place DVM probes into rear sockets volts dc with power off and 7 2 7 5 volts dc with power on 4 Using a DVM measure the excitation voltage EXE and EXE in Figures 3 2 and 3 3 The readin...

Страница 25: ...oad Cell assembly non standard proprietary connection Figure 3 3 shows the difference between standard and non standard strain gauge load cells The wires on the standard strain gauge load cell left co...

Страница 26: ...receives a voltage excitation from a regulated power source in the Indicator and returns a low voltage millivolt mV signal relative to the force applied The Load Cell signal is converted to a numeric...

Страница 27: ...nted Circuit Board PCB Figure 3 11 Load Cell wiring 3 3 2 3 How to Operate Simulators The function of a Simulator is to simulate the output of a full bridge Load Cell allowing the scale to be separate...

Страница 28: ...dicator housing Also remove the battery 2 Disconnect the Load Cell cable from the PC Board 3 With the Simulator set to zero attach it to the PCB See Figure 3 11 4 Connect a known good AC Adapter to th...

Страница 29: ...gures 3 2 and 3 3 5 Carefully disconnect the AC power and battery connectors and the Membrane Switch cable 6 Remove the four screws from the PCB 7 Lift out the PCB CAUTION When handling the PCB grasp...

Страница 30: ...the Top Cover and remove all traces of adhesive 3 Remove the protective backing from the back of the new Function Label Carefully position it on the Top Housing starting at the bottom of the cover Us...

Страница 31: ...he software revision and then goes into weighing mode 4 If all of the displays appear normal in this test continue with the next test If the display did not appear check the AC adapter output and repe...

Страница 32: ...r to each setting between 0 2 mv V and 1 8 mv V The indicator reading should be stable and linear throughout the range of settings 4 5 Calibration Retention Test This test checks that the Indicator re...

Страница 33: ...r in the field NOTE In all cases where a part is replaced the indicator must be thoroughly checked after the replacement is made The indicator MUST meet the parameters of all applicable specifications...

Страница 34: ...ISTS DIAGRAMS Ohaus Corporation www ohaus com 5 2 3000 Series T32ME MC Indicators Service Manual 5 1 3000 SERIES T32MC LCD INDICATOR HOUSING INTERNAL PARTS Figure 5 1 Series 3000 T32MC Indicator Housi...

Страница 35: ...OR HOUSING INTERNAL PARTS TABLE 5 1 T32MC LCD INDICATOR HOUSING INTERNAL PARTS Drawing Item Description 1 LFT Kit 2 Function Label 3 Housing Top 5 LCD W backlight 7 Main PCB 10 Bottom Housing 11 Knobs...

Страница 36: ...LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 4 3000 Series T32ME MC Indicators Service Manual 5 2 3000 SERIES T32ME INDICATOR HOUSING INTERNAL PARTS Figure 5 1 Series 3000 T32ME Indicator Housing...

Страница 37: ...ICATOR HOUSING INTERNAL PARTS TABLE 5 2 T32ME LED INDICATOR HOUSING INTERNAL PARTS Drawing Item Description 1 LFT Kit 2 Function Label 3 Housing Top 5 LED 7 Main PCB 10 Bottom Housing 11 Knobs set of...

Страница 38: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 6 3000 Series T32ME MC Indicators Service Manual...

Страница 39: ...again after turning it off it is necessary to have the Indicator locked and sealed by local Weights and Measures authorities 4 If the Calibration Menu is locked in the Menu Lock submenu turn it back o...

Страница 40: ...pan calibration was successful the scale exits to the active weighing mode and displays the actual weight value A 3 LINEARITY CALIBRATION Linearity calibration uses three calibration points the first...

Страница 41: ...0 1 5 10 15 20 25 30 40 50 60 75 100 50 100 120 0 01 0 02 0 05 0 1 0 02 0 05 0 1 5 10 15 20 25 30 40 50 60 75 100 120 60 120 150 0 01 0 02 0 05 0 1 0 05 0 1 5 10 15 20 25 30 40 50 60 75 100 120 150 7...

Страница 42: ...2000 2500 3000 5000 2500 5000 6000 0 5 1 2 5 1 2 5 5 10 15 20 25 30 40 50 60 75 100 120 150 200 250 300 400 500 600 750 1000 1200 1500 2000 2500 3000 5000 6000 2500 5000 7500 0 5 1 2 5 2 5 5 10 15 20...

Страница 43: ...JUSTMENT VALUES Elevation above sea level in meters 000 325 325 650 650 975 0975 1300 1300 1625 1625 1950 1950 2275 2275 2600 2600 2925 2925 3250 3250 3575 Elevation above sea level in feet Geographic...

Страница 44: ...85 45 85 45 90 00 15 15 16 16 17 17 18 18 19 19 20 20 21 21 22 22 23 23 24 24 25 25 26 26 27 27 28 28 29 29 30 30 31 14 15 15 16 16 17 17 18 18 19 19 20 20 21 21 22 22 23 23 24 24 25 25 26 26 27 27 28...

Страница 45: ......

Страница 46: ...80253153 P N 80253153 SERVICE MANUAL 3000 SERIES T32MC T32ME INDICATORS...