![OHAUS Ranger RC Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/ohaus/ranger-rc/ranger-rc_instruction-manual_1633084040.webp)
5.3 TOLEDO Continuous mode
5.3.1
TOLEDO Continuous commands
The scale supports the following input commands in TOLEDO Continuous mode:
5.3.2
Output format in TOLEDO Continuous mode
Weight values are always sent in the following format in TOLEDO Continuous mode:
Command
Meaning
P
<CR><LF>
Print out the current result
T
<CR><LF>
Tare the scale
Z
<CR><LF>
Zero the display
C
<CR><LF>
Clear the current value
T
x.xxx <CR><LF>
Define tare
Status
Field 1
Field 2
1
2
3
4
5
6
7
8
9
10
11
12 13 14 15 16
17
18
STX SWA SWB SWC MSD –
–
–
–
LSD MSD –
–
–
–
LSD CR
CHK
Field 1
Field 2
STX
SWA, SWB, SWC
MSD
LSD
CR
CHK
6 digits for the weight value that is sent without a decimal point and unit
6 digits for the tare weight that is sent without a decimal point and unit
ASCII characters 02 hex, characters for "start of text"
Status words A, B, C, see below
Most significant digit
Least significant digit
Carriage Return, ASCII characters 0D hex
Checksum (2-part complement of the binary sum of the 7 lower bits of all previously sent
characters, incl. STX and CR)
Ranger RC
Interface description