![OHAUS Ranger RC Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/ohaus/ranger-rc/ranger-rc_instruction-manual_1633084037.webp)
5.2 SICS interface commands
The Ranger RC compact scales support the command set MT-SICS (METTLER
TOLEDO
S
tandard
I
nterface
C
ommand
S
et). With SICS commands, it is possible to
configure, query and operate the scales from a PC. SICS commands are divided up
into various levels.
5.2.1
Available SICS commands
Command
Meaning
LEVEL 0
@
Reset the scale
I0
Inquiry of all available SICS commands
I1
Inquiry of SICS level and SICS versions
I2
Inquiry of scale data
I3
Inquiry of scale software version
I4
Inquiry of serial number
S
Send stable weight value
SI
Send weight value immediately
SIR
Send weight value repeatedly
Z
Zero the scale
ZI
Zero immediately
LEVEL 1
D
Write text into display
DW
Weight display
K
Keyboard check
SR
Send and repeat stable weight value
T
Tare
TA
Tare value
TAC
Clear tare
TI
Tare immediately
LEVEL 2
C2
Calibrate with the external calibration weight
C3
Calibrate with the internal calibration weight
I10
Inquire or set scale ID
I11
Inquiry of scale type
P100
Print out on the printer
Interface description
Ranger RC
37
To use the MT-SICS commands, first send the OHAUS command PSI. To return to the
OHAUS command set, send the OHAUS command POH.