Ranger™ 2000
Series EN-9
5. The scale will display [
CL.lImM
] (clear Check limits).
6. Press
No
to use the stored check limits and proceed to step 9.
7. Press
Yes
to establish new check limits. The scale will then display [
Set. Lo
].
Press
Yes
to view the “Low” limit value. Press
Yes
to accept or
No
to edit
the “Low” limit value. The stored value then displays with the first digit
highlighted [
0
00.000
%]. Repeatedly press
No
until the desired number
appears. Press
Yes
to accept and highlight the next digit.
Repeat until all the digits are correct. Press
Yes
to accept the “low” limit
value, [
Set. Hi
] will be displayed.
8.
Repeat the same procedure to accept or edit the “high” value.
9. If required, place an empty container on the pan and press
Tare
.
10. Place sample material on the pan or in the container. If the sample weight is
under the target percentage range, the yellow LED will light. If the sample is
within the target percentage range, the green LED will light. If the sample is
over the target percentage range, the red LED will light.
Note:
Press
Function
to view the low and high check limits.
For numeric settings, please see section 4.2 for details.
3.6.3 Check Count
Set Checkweigh to Count in the Mode menu. Use this mode to compare the
quantity of items to a target quantity range.
1. Press and hold
Mode
until [
CHeCk
] (Check) is displayed. [
CLr.PwW
] (clear
APW) will then display.
If no APW exists, the scale will display [
Put.10
], proceed to step 5.
2. Press
No
to use the stored APW and proceed to step 5.
3. Press
Yes
to establish a new APW. Scale will now display [
Put.10
]. Press
No
or
Back
to toggle the choices (5, 10, 20, 50 or 100).
4. Put the indicated number of pieces on the pan then press
Yes
to calculate
the APW.
5. The scale will display [
CL.lImM
] (clear Check limits).
6. Press
No
to use the stored check limits and proceed to step 9.
7. Press
Yes
to establish new check limits. The scale will then display [
Set. Lo
].
Press
Yes
to view the “Low” limit value. Press
Yes
to accept or
No
to edit
the “Low” limit value. The stored value then displays with the first digit
highlighted [
0
00000
Pcs]. Repeatedly press
No
until the desired number
appears. Press
Yes
to accept and highlight the next digit.
Repeat until all the digits are correct. Press
Yes
to accept the “low” limit
value, [
Set. Hi
] will be displayed.
8.
Repeat the same procedure to accept or edit the “high” value.
9. If required, place an empty container on the pan and press