EN-21
3000 Series Indicators
3.4.7 Zero Range
Set the percentage of scale capacity that may be zeroed.
0%
= zeroing disabled
2% = zero up to 2 percent of capacity
100% = zero up to full capacity
•
•
•
3.4.6 Power On Unit
Set the unit that will be active at power on.
oz, lb, g, kg, lb:oz or
Auto (last unit in use when power was turned off.)
3.4.8 End Setup
Advance to the next menu.
3.5 Readout Menu
Enter this menu to customize display functionality.
Reset:
No
, Yes
Filter Level Lo,
Med
, Hi
Auto Zero Tracking Off,
0.5d
, 1d, 3d
Backlight Off, On,
Auto
Auto Shut Off
Off
End Readout
Exit READOUT menu
3.5.1 Reset
Set the Readout menu to factory default settings.
No
= not reset
Yes = reset
If the Legal for Trade menu item is set to ON, the Stable Range, Averaging Level, Auto Zero Tracking and Auto
Off settings are not reset.
3.4.5 Graduation
Set the scale readability.
0.001, 0.002, 0.005, 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1, 2, 5, 10, 20.
NOTE
: Not all settings are available for each capacity. Refer to the Setup Table 3.3 for available settings.
Содержание Ranger 3000 Series
Страница 1: ...i Indicatori serie 3000 Manuale di istruzioni Indicatore T31XW Indicatore T31P...
Страница 2: ...ii...
Страница 45: ...i Manuel d instructions pour les indicateurs de la s rie 3000 Indicateur T31XW Indicateur T31P...
Страница 46: ...ii...
Страница 89: ...i Indicadores serie 3000 Manual de instrucciones Indicador T31XW Indicador T31P...
Страница 90: ...ii...
Страница 133: ...3000 Series Indicators Instruction Manual T31P Indicator T31XW Indicator...
Страница 134: ......
Страница 179: ...i Bedienungsanleitung f r Indikatoren der Serie 3000 T31XW Indikator T31P Indikator...
Страница 180: ...ii...