APPENDIX A - USER CALIBRATIONS
Ohaus Corporation www.ohaus.com
38
Pioneer PJX Series Service Manual
5. Once the span calibration is completed successfully,
CAL.Done
will display.
1.1
Press any button to return to the previous
screen.
6. Remove the weight(s), and the reading will be set to
zero.
A.3
LINEARITY CALIBRATION
Linearity calibration uses three calibration points, one at zero load and the others at specified loads.
With no load on the scale, press Linearity Calibration to begin the process.
The balance captures the zero point, and then prompts for the next weight.
Continue to follow the instructions on the display until the calibration is completed.
Steps for linearity calibration
1. Long press the
Tare / Menu-Cal
button, and the
Calibration Menu will display.
2. Press
Yes
to enter the Calibration Menu.