CHAPTER 2 - DIAGNOSTIC GUIDE
Ohaus Corporation www.ohaus.com
12
Pioneer PJX Series Service Manual
TABLE 2-1
DIAGNOSTIC GUIDE
Symptom / Display
Possible Cause
Err 8.1
Weight reading exceeds Power On Zero limit.
Err 8.2
Weight reading below Power On Zero limit.
Err 8.3
Weight reading exceeds overload limit.
Err 8.4
Weight reading below underload limit.
Ref Err
Reference Weight too small. The weight on the pan is too small to define a
valid reference weight. Add additional samples.
LowW.rEF
A warning that the reference weight is below the recommended value for
accurate parts counting. Add additional samples if possible otherwise
counting continues with diminished accuracy.
-------
Busy (tare, zero, and printing).
--
NO
--
Function not executed.
Error
Timeout
IDnr.Err
IDNR Error