background image

EXPLORER BALANCES 

 

EN-37 

 

 

Accumulating displayed values & Viewing data 

– Manual 

Once Statistics is enabled, two buttons appear on the Application screen:39 
 

Statistics

 and 

Accumulate

 

To add the sample to the data set, place the item on the pan and 
press 

Accumulate

.  

When stable, the Statistics button lights briefly to show the item 
is added to the set and the data set quantity increments by one. 

 

 

 

Continue to add items and press 

Accumulate

 to build the 

statistical data set  

At any time, press the 

Statistics

 button to view the results.

 

 

 

 

The 

Statistics Results

 screen 

Press 

Print

 to send to the enabled COM port 

Press 

Close

 to return to the Application home screen with 

the current statistics set or 

Press 

Clear Data

 to clear the results and return to the 

Application home screen.

 

 

 

 

Statistics 

– Automatic 

Statistics Automatic mode is similar to manual mode (above) except the first stable weight is automatically 
captured. Thus the Accumulate button is not used. 
 

www.

.com

[email protected]

1.800.561.8187

Содержание Explorer EX35001N

Страница 1: ...Explorer Balances Instruction Manual www com information itm com 1 800 561 8187...

Страница 2: ...www com information itm com 1 800 561 8187...

Страница 3: ...Home Screen 15 3 2 Principal Functions and Main Menu 15 3 3 Overview of Parts and Features 16 3 4 Overview of Parts and Features Draft Shield Models 17 3 5 Overview of Parts and Features Non Draft Shi...

Страница 4: ...ential Weighing 64 4 9 1 Item Settings 64 4 9 2 Application Setup 66 4 9 3 Differential Weighing with Auto Process On 67 4 10 Density Determination 68 4 10 1 Measuring the Density of a Solid Using Wat...

Страница 5: ...tings 104 5 5 6 Functions Configuration 105 5 5 7 TouchLess Sensor Setup 105 5 5 8 Draft Shield light 105 5 5 9 Ionizer seconds 105 5 6 Balance Setup 106 5 6 1 Balance Set up sub menu 106 5 6 2 Stable...

Страница 6: ...and GMP Data 120 5 13 8 Reset Communication 120 5 13 9 Reset All Global Reset 120 5 14 Securing the Menu and Key Lock menu settings 120 5 15 Information 121 5 15 1 Using Information 121 6 LEGAL FOR T...

Страница 7: ...ing Counting Percent Weighing Check Weighing Dynamic Weighing Filling Ionizer EX225D AD and EX225 AD only The Ionizer generates bipolar ions continuously by DC corona discharge separately from positiv...

Страница 8: ...tform Dry off any liquid spills immediately The instrument is not watertight Do not immerse the equipment in water or other liquids Only connect the power cord to a compatible grounded electrical outl...

Страница 9: ...10201 EX10202 EX12001 EX24001 EX35001 Terminal X X X X X Weighing Base X X X X X Weighing Platform 80 mm Diameter X 90 mm Diameter X 130 mm Diameter X 190 x 200 mm X 311 x 377 mm X Sub Platform 80 mm...

Страница 10: ...identical for all Explorer Balance models 2 2 2 Installing the Wind Ring Weighing Platform on Semi Micro models Note Assembly instructions also apply to models with the suffix M or N 1 Place the Wind...

Страница 11: ...g side doors be sure to set the lock tabs in the proper position as shown here UNLOCK LOCK 2 Follow the steps below to install side front and top doors of the Draft Shield Side Door Installation 1 Unl...

Страница 12: ...glass doors on both sides and open the top lid to make space for the glass shelf 2 2 8 Installing the Weighing Pan Wind Shield Non Draft Shield Models Note Assembly instructions also apply to models...

Страница 13: ...otate The Thumbwheels must be used for leveling See the Level Assist screen in the Information Menu To see the Level Assist menu touch More on the bottom right corner of the main screen Then touch Lev...

Страница 14: ...to connect either to a computer or a printer with a standard straight through serial cable Or connect using the scale s USB ports Note For configuration and interface commands see the Communication M...

Страница 15: ...the Base A cable is attached to the Terminal Take care to not damage or disconnect this cable 2 To reattach the Terminal press in the two Release buttons and slide the Terminal into the Base until the...

Страница 16: ...plorer Balances have built in AutoCal which can calibrate the balance automatically and does not require calibration masses If preferred the balance can be manually calibrated with external masses Hav...

Страница 17: ...e display to zero Place an item on the pan Display indicates gross weight Taring With no load on the pan press Tare to set the display to zero Place an empty container on the pan and press Tare Add ma...

Страница 18: ...o Balance Leveling Thumbwheels Data label Terminal cable storage clips Lock switch Weigh Below Access Foot Non Adjustable Type label Far side M N models only Touchless Sensors Power Connection Draft S...

Страница 19: ...xplorer Base Leveling Thumbwheels Data label Terminal cable storage clips Lock switch Weigh Below Access Foot Non Adjustable Type label Far side M N models only Touchless Sensors Power Connection Draf...

Страница 20: ...ew Figure 3 3 Explorer Base Illuminated Level Indicator Touchless Sensors Touchless Sensors with Indicator Lights Type label Far side M N models only Data label Terminal cable storage clips Lock switc...

Страница 21: ...01 and EX35001 Top View Bottom View Rear View Side View Figure 3 4 Explorer Base Touchless Sensors Level Indicator Power Connection Leveling Thumbwheels Foot Non Adjustable Terminal cable storage clip...

Страница 22: ...rminal Hooks to attach to base Capacity sticker Approved models only Touchless Sensors with Indicator Lights Terminal Bracket Terminal Release Buttons Angular Adjust Buttons Foot USB Device Connection...

Страница 23: ...sensors that can be assigned a unique function when activated e g zero print ionizer tare Automatic draftshield door etc To assign a task for each TouchLess sensor press Sensors Note To activate a sen...

Страница 24: ...g Unit can also be accessed by touching the unit area of the Main Display line If a unit is not displayed in the list it must first be turned on via the User Menu 3 8 14 Ionizer only for EX225D AD and...

Страница 25: ...urrent Application 3 8 19 Logout Used for User logout and return to User Login screen This feature is only active if a user profile has been created and a User has logged in 3 9 Ionizer EX225D AD and...

Страница 26: ...touch Enabled in the setup menu for that application mode Percent Weighing in the example below Each enabled mode appears in the Applications menu The Explorer Balance incorporates the following Appli...

Страница 27: ...of measure The WEIGHING Home screen Main Display Line Second Display Line not shown Reference Fields Application Buttons Functions Application Icon 4 1 1 Edit Settings To view or adjust the current s...

Страница 28: ...to Application at bottom right of that screen To input a sample name touch the Sample Name button Enter the desired sample name and then press You can use bar code scanner to input sample name batch I...

Страница 29: ...Application on off RFID On Off To read RFID code to get sample name Secondary Unit On Off To show the Second Display line Minimum Weight On Off To enable Min Weight Statistics Off Manual Auto Lock Set...

Страница 30: ...for Auto Save to USB Auto Save to USB applies to Weighing Parts Counting Percent Weighing Check Weighing and Filling 4 1 5 Statistics for Weighing The Statistics function allows you to compare a numb...

Страница 31: ...nt sample must be removed before capturing the next sample When stable the Statistics button lights briefly to show the item is added to the set and the data set quantity increments by one Continue to...

Страница 32: ...Edit Settings To view or adjust the current settings touch the Edit Settings button The Settings screen is now displayed Settings available APW Sample Size Sample Name Sample ID Batch ID Clear All Not...

Страница 33: ...o establish a new APW touch the Recalculate APW button The APW Screen is displayed Note The sample size that is displayed will be used To use a different sample size change that first See below Follow...

Страница 34: ...ouch the Samples button A numeric input window appears Key in the desired Sample Size then press The next screen appears with the message to place the number of pieces entered in the numeric input win...

Страница 35: ...Off To show the Second Display line Statistics Off Manual Auto Lock Settings On Off Press to lock the current setting s so they cannot be changed Reference fields APW On Off The APW is displayed Gross...

Страница 36: ...set The Average Piece Weight and standard sample size can also be set Follow these steps Touch the Edit Settings button The Settings screen appears The buttons for APW Samples Sample Name Sample ID Ba...

Страница 37: ...he PARTS COUNTING Fill Home screen Main Display Line Second Display Line Reference Fields Application Buttons Functions The default or last Average Piece Weight APW is displayed Place objects on the p...

Страница 38: ...ples and examine the relative deviation of the samples along with other statistical data A minimum of three samples is required Statistics can be determined manually Manual mode with key press after e...

Страница 39: ...nue to add items and press Accumulate to build the statistical data set At any time press the Statistics button to view the results The Statistics Results screen Press Print to send to the enabled COM...

Страница 40: ...isplay Line Reference Fields Application Buttons Functions Application Icon 4 3 1 Edit Settings To view or adjust the current settings To adjust the Application settings touch the Edit Settings button...

Страница 41: ...the Recalculate Reference button The New Reference screen appears Follow the screen instructions to establish a new Reference Weight Actions available Tare Cancel or Accept Press Accept to capture th...

Страница 42: ...creen press Done Percent Weighing Application Setups defaults in Bold Item Available Settings Comments Enabled On Off Turns Application on off Secondary Unit On Off To show the Second Display line Loc...

Страница 43: ...plication Icon Functions 4 4 2 Edit Settings To view or adjust the current settings Touch the Edit Settings button The Settings screen appears The Application Settings Screen Settings available Over L...

Страница 44: ...reading is under or over Statistics Off Manual Auto Display Setting Weight Check Status To show Weight or Check Status on Main Display Line Lock Settings On Off Press to lock the current setting s so...

Страница 45: ...to begin Check Weighing CHECK WEIGHING Nominal Wt Tolerance home screen Main Display Line Second Display Line not shown Reference Fields Application Buttons Functions Application Icon To use this mod...

Страница 46: ...n Check Weighing The CHECK WEIGHING Nominal Percent Tolerance home screen Main Display Line Second Display Line not shown Reference Fields Application Buttons Application Icon Functions To use this mo...

Страница 47: ...eighing Line set the Display Setting to Check Status using the Application Setup screen The default value is Weight The CHECK WEIGHING home screen with Display set to Check Status Main Display Line Se...

Страница 48: ...two buttons appear on the Application screen Statistics and Accumulate To add the sample to the data set place the item on the pan and press Accumulate When stable the Statistics button momentarily li...

Страница 49: ...fault or last Averaging Time is displayed Follow the screen instructions The DYNAMIC WEIGHING Home screen Main Display Line Second Display Line Reference Fields Application Buttons Functions Applicati...

Страница 50: ...et from 1 to 99 seconds The default value is 10 To change it press the button A numeric input window is displayed Key in the desired Averaging Time then press Done The display returns to the previous...

Страница 51: ...inting Section Averaging Time On Off Auto Print On Off To print automatically after weighing process is finished 4 5 3 Dynamic Weighing Semi Auto The cycle begins when an object animal is placed on th...

Страница 52: ...delay To enable this feature enter the Dynamic Weighing Setup menu Press Auto Sample and change the mode to Automatic The Dynamic Automatic screen appears with the default or last Averaging Time displ...

Страница 53: ...tatistics is enabled two buttons appear on the Application screen Statistics and Accumulate To add the sample to the data set place the item on the pan and press Accumulate When stable the Statistics...

Страница 54: ...tings screen appears Settings available Target Weight Sample Name Sample ID Batch ID Clear All Note please refer to section 4 1 1 for details about setting Sample Name Sample ID Batch ID and Clear All...

Страница 55: ...to the Settings screen The display returns to the Settings screen with the new Target Weight value highlighted momentarily To return to the Filling home screen touch Return to Application 4 6 2 Applic...

Страница 56: ...Off Tare weight is displayed Print Options See Printing Section Target Weight On Off Difference Weight On Off Difference Percent On Off 4 6 3 Statistics for Filling Use the Statistics function to com...

Страница 57: ...nue adding items and pressing Accumulate to build the statistical data set At any time press the Statistics button to view the results The Statistics Results screen Press Print to send to the enabled...

Страница 58: ...ication Modes scroll down to find Totalization press it then press Enabled to turn it On It will now appear in the Application Modes menu on the home screen Place objects on the pan to begin The TOTAL...

Страница 59: ...uto Sample Mode On Off To select the mode Reference fields Samples On Off To display the number of samples weighed Average weight On Off Displays the Average Standard deviation On Off Displays the sta...

Страница 60: ...r the pan if necessary Add a new sample The item weight is shown on the Main Display Line The weight of the item is automatically added when stable to the total The new total is shown on the Second Di...

Страница 61: ...the first ingredient on the pan or in a tared container Press Accept to store the first component A prompt says Place item 2 Press Accept Continue adding components and pressing Accept to store the we...

Страница 62: ...e Formulate Recipe Based To select the mode Filler On Off To prompt for a filler material Secondary Unit On Off To show the Second Display line Reference fields Total On Off Total is displayed Gross O...

Страница 63: ...PE FORMULATION Home screen Main Display Line Second Display Line Reference Fields Application Buttons Functions Application Icon Press Begin Formulation If no items have been previously set up the Set...

Страница 64: ...the Main Display Line The progress bar displays the fill status Within 10 percent of the target value the progress bar converts to fine resolution display Press Accept to add the weight of the compon...

Страница 65: ...ANCES EN 63 Formulation Results Press Print to send the results to the enabled COM port To close the Results window and return to the Application home screen press Close www com information itm com 1...

Страница 66: ...weights are gathered separately for each item after each item is selected in Item Settings Touch the Setup icon wrench to enter Application Setup from the home screen The DIFFERENTIAL WEIGHING Home sc...

Страница 67: ...This feature is disabled if no final weight is stored To clear all items press Clear All then confirm When finished press Return to Application On the main Application screen follow the screen instruc...

Страница 68: ...s Initial weight The Differential Weighing Application Setups defaults in Bold Item Available Settings Comments Enabled On Off Turns Application on off Auto Process Mode On Off To select the mode Refe...

Страница 69: ...Display Line Second Display Line not shown Reference Fields Application Buttons Functions Application Icon Follow the screen instructions Press Item Settings to set up items in memory if necessary Us...

Страница 70: ...key press after each step or automatically weights are automatically recorded Before making density measurements establish the Application Settings The DENSITY DETERMINATION Home screen Main Display...

Страница 71: ...below Step 1 of 2 Weigh the Sample in Air Press Begin Density Calculation Follow screen instructions then press Accept to store the dry sample weight in air Step 2 of 2 Weigh the Sample Submerged in...

Страница 72: ...ents Porous Material On Off To select Porous material Auto Print Result On Off To automatically print results Auto Sample Mode On Off To select the mode Reference fields Liquid Density On Off Test liq...

Страница 73: ...plication Setup The DENSITY DETERMINATION SOLID Home screen Main Display Line Second Display Line Reference Fields Application Buttons Functions Application Icon Confirm the default values displayed L...

Страница 74: ...ns Application Icon Confirm that the default value displayed sinker volume is correct To edit the default values touch Edit Settings and the Settings screen is now displayed Settings available Sinker...

Страница 75: ...Follow the screen instructions then press Accept to store the sinker weight submerged in liquid Once the necessary weights have been determined the density of the Liquid sample is displayed in g cm3 a...

Страница 76: ...er Temp Oil Density Sample Name Sample ID Batch ID Clear All Note please refer to section 4 1 1 for details about setting Sample Name Sample ID Batch ID and Clear All Functions Available Return to App...

Страница 77: ...Oiled Sample Submerged in Liquid Follow the screen instructions then press Accept to store the oiled sample weight submerged in liquid Once the necessary weights have been determined the density of t...

Страница 78: ...stable and unstable weights to be captured 4 11 1 Peak Hold Semi Automatic default In the upper left portion of the home screen select Peak Hold Follow the screen instructions to begin The PEAK HOLD H...

Страница 79: ...old Application Setups defaults in Bold Item Available Settings Comments Enabled On Off Turns Application on off Auto Sample Mode Manual Semi Automatic Automatic To select the mode of operation Stable...

Страница 80: ...Display Line Second Display Line Reference Fields Application Buttons Functions Application Icon Follow the screen instructions Place item on the pan Press Begin Peak Hold to begin the capture process...

Страница 81: ...POINT Auto Home screen Main Display Line Second Display Line Reference Fields Application Buttons Functions Application Icon Place item on the pan The capture process starts automatically The highest...

Страница 82: ...ference Fields Application Buttons Functions Application Icon 4 12 1 Item Settings Press Item Settings to set up items in memory The Edit Settings Screen Settings available Add New Edit name Edit Cost...

Страница 83: ...highlighted momentarily To change the default component Unit press Edit Unit The Weighing Units screen appears Continue adding items as required using Add New New items can be added at any time To del...

Страница 84: ...the screen instructions to add the items to the pan and touch the Accept button until each item has been weighed Press the View Data button at any time to show the current results Item Item Cost and I...

Страница 85: ...osting Application Setups defaults in Bold Item Available Settings Comments Enabled On Off Turns Application on off Currency label None various To select the Currency label Reference fields Unit Cost...

Страница 86: ...p to 30 samples can be used The PIPETTE ADJUSTMENT Home screen Main Display Line Second Display Line Reference Fields Application Buttons Functions Application Icon 4 13 1 Pipette Adjustment Manual de...

Страница 87: ...to the previous screen with the new value highlighted Continue to enter the required information following the above procedure Pipette Inaccuracy via the Inaccuracy button Barometric Pressure via the...

Страница 88: ...pt to store each sample weight liquid The default number of samples is 10 It can be changed in Setup from 2 to 30 samples Once all the sample weights have been dispensed the Results screen is automati...

Страница 89: ...during measurements Unit of Pressure ATM inHg KPa mbar mmHg PSIA To select Atmospheric Pressure unit Volume Unit Pipette capacity unit ml l Nominal capacity unit Secondary Unit On Off To show the Sec...

Страница 90: ...C collection can be done manually Manual mode requires key press after each step or automatically Auto Mode the weights are automatically recorded when stable In the upper left portion of the home scr...

Страница 91: ...hen displayed Follow the screen instructions then press Accept to store the sample weight The current number of Samples along with the Average Sum Minimum value Maximum value and Range are displayed C...

Страница 92: ...turn to the SQC home screen Press Close to return to the SQC home screen to begin another batch or to view process results To view the Process results graphically batch comparison press the View Trace...

Страница 93: ...Off Press to lock the current setting s so they cannot be changed Reference fields Number of Batches On Off The total batches Number of Samples On Off The total samples Average On Off The Average valu...

Страница 94: ...les can be tested Each sample is shown with weight difference compared to the average weight and if it is accepted or failed In the upper left portion of the home screen select Fill Weight Variation T...

Страница 95: ...ss The default value is 20 pieces A new screen appears with the message Place samples to average Press Accept to continue Place all samples together on the pan to calculate the sample average weight T...

Страница 96: ...ch the Accept button to store each sample s weight The maximum sample number is 99 It can be changed from 1 to 99 Once all the samples have been weighed the Results screen appears automatically Setup...

Страница 97: ...changed Reference Difference Weight On Off Weight difference is displayed Difference Percent On Off Weight difference percentage is displayed Average On Off Weight average is displayed Limit On Off L...

Страница 98: ...x x x Totalization Formulation x x Differential Weighing Density Determination x x x x x Peak Hold Ingredient Costing Pipette Adjustment x x SQC x x Fill Weight Variation 4 17 1 Creating a Library Rec...

Страница 99: ...t press Delete instead of Recall 4 18 Additional Features 4 18 1 Weigh Below The Explorer Balance is equipped with a weigh below hook for weighing below the balance Note before turning the balance ove...

Страница 100: ...ain Menu List and touch it The Sub Menu appears Select the Menu Item Continue until the desired setting is chosen in the Menu list Touch the setting to change it The changed setting will be displayed...

Страница 101: ...temperature and is level With the Balance turned ON and no load on the pan touch AutoCal Internal Calibration and then touch Calibration The Balance begins to calibrate Alternately touch the shortcut...

Страница 102: ...5 1 below Example Actual weight reading 200 014 Expected weight reading 200 000 Test mass value Difference Weight d 0 014 Difference weight in digits 14 Adjust value To perform a Calibration Adjustme...

Страница 103: ...e display The best accuracy is achieved using the mass closest to the full span value Note To change the span calibration point touch the alternate weight shown on the display Follow the screen instru...

Страница 104: ...New button and insert a user name and then press To edit a user select the user you want to edit and press the Edit button You can change User Name Password and Group for each user To edit the passwor...

Страница 105: ...ration setting x x External calibration x Internal calibration x Date Time x x x Application modes settings x x Units x x Device Print settings x x GLP x x Balance name x x x library x x System log x...

Страница 106: ...10 characters with combination of letters and numbers 5 5 3 Language Set the language displayed for menus and displayed messages English German French Spanish Italian Chinese Korean Japanese Russian P...

Страница 107: ...Left Default setting for Automatic draftshield door models Terminal left left door Terminal right right door Note EX12001 EX24001 and EX35001 do not have Base sensors 5 5 8 Draft Shield light Set the...

Страница 108: ...ter stabilization time with less stability MEDIUM normal stabilization time with normal stability HIGH slower stabilization time with more stability The setting is locked at the current setting when t...

Страница 109: ...M internal Calibration menu will be locked Balance Setup Menu Filter Level is locked at the current setting Stability Indicator Range is locked at 1 Division Auto Zero Tracking is limited to 0 5 Divis...

Страница 110: ...108 EXPLORER BALANCES 5 6 10 Date Time Set the current Date and Time Change the format if desired then enter the current value Select to return to the sub menu www com information itm com 1 800 561 81...

Страница 111: ...tions shortcut top left of all home screens The current menu item status is shown OFF disabled ON enabled 5 8 Weighing Units Enter this sub menu to activate the desired units of measure Note Due to na...

Страница 112: ...s set to ON the Factor Exponent and Least Significant Digit must be set Factor Set the conversion factor using the numeric keypad Settings of 0 00001 to 1 9999999 are available The default setting is...

Страница 113: ...ble Alphanumeric settings up to 25 characters are available for each Header setting 5 9 4 Balance Name Set the balance identification Alphanumeric settings up to 25 characters are available The defaul...

Страница 114: ...ions Mini B Use this interface to connect the balance to a computer or printer Choosing an item brings up another menu level RS232 shown Device Settings Print Settings Choosing a Settings item brings...

Страница 115: ...tings Menu Note If LEGAL FOR TRADE is set to ON the print Output Stable setting is not reset 5 10 6 Print Output Print Stable Weight Only Set the printing criteria OFF values are printed immediately r...

Страница 116: ...ue is printed Balance Id XXXXXXX Balance Name Set the status OFF the Balance Name is not printed ON the Balance Name is printed Balance Name XXXXXXX User Name Set the status OFF the User Name is not p...

Страница 117: ...ion Set the status OFF the reference info is not printed ON the reference info is printed Note See below for more clarification Signature Line Set the status OFF the Signature Line is not printed ON t...

Страница 118: ...kg Max 1 001 kg Diff 0 002 kg Counting APW 0 100 kg N 10 Total 1000 Pcs Avg 100 Pcs Std 1 Pcs Min 99 Pcs Max 101 Pcs Diff 2 Pcs Check Weighing Under 0 995 kg Over 1 005 kg Under 0 995 kg Over 1 005 k...

Страница 119: ...Lines move the paper up four lines after printing Form a form feed is appended to the output 5 10 11 Data Transfer Function Output weighing results directly to a PC application Setup is easy and no ad...

Страница 120: ...Otherwise Excel will not distinguish it as a negative number 5 11 Library When an item is processed on a regular basis the item s data may be stored in memory for future use This memory is referred t...

Страница 121: ...r To verify proper operation of each Touchless Sensor When operated each sensor should light and make a sound Note Only for Automatic draftshield door models 5 12 6 Data Maintenance To check Software...

Страница 122: ...YES to return all GLP and GMP Data menu items to their factory settings 5 13 8 Reset Communication Select YES to return all Communication menu items to their factory settings 5 13 9 Reset All Global R...

Страница 123: ...rtiption of the item selected is displayed This feature is menu based 5 15 1 Using Information Touch the Info Icon to bring up the menu UPPER RIGHT Information topics beginning with Level Assist Touch...

Страница 124: ...ained in Section 5 3 Set the position of the Security Switch to the locked position 4 Set Legal for Trade to ON in the Balance Setup menu 6 2 Verification A weights and measures official or authorized...

Страница 125: ...erties Settings then ASCII Setup Check boxes as illustrated Send line ends Echo typed characters Wrap lines Use RS232 Interface Commands Section 9 6 1 to control the balance from a PC Note When the Hy...

Страница 126: ...R User Name Project Name Application Weighing Sample Name Batch ID Sample ID 2 771 g Gross 4 148 g Net 2 771 g N Tare 1 377 g T Minimum Weight 0 000 g Signature Verified by Company Name Address 1 Addr...

Страница 127: ...Under Limit 10 PCS Signature Verified by Parts Counting Fill Parts Counting Statistics Company Name Address 1 Address 2 Address 3 Address 4 12 7 2010 3 23 00 PM Balance ID Balance Name EXPLORER User...

Страница 128: ...e ID Balance Name EXPLORER User Name Project Name Project Name Application Check Weighing Sample Name Sample ID Batch ID Result Under Gross 1 401 g Net 0 455 g N Tare 0 946 g T Target 310 000 g wt tol...

Страница 129: ...ol 50 00 Percent Tol 50 00 Signature Verified by Company Name Address 1 Address 2 Address 3 Address 4 12 7 2010 3 37 00 PM Balance ID Balance Name EXPLORER User Name Project Name Sample Name Sample ID...

Страница 130: ...0 g RSDEV 0 0 Minimum 0 03 g Maximum 0 03 g Range 0 00 g Signature Verified by Filling Filling Statistics Company Name Address 1 Address 2 Address 3 Address 4 12 7 2010 3 42 00 PM Balance ID Balance N...

Страница 131: ...n Formulation Sample Name Sample ID Batch ID Total Target 3 000 g Total Actual 5 401 g Total Diff 80 0 Start Time 12 7 2010 3 35 00 PM End Time 12 7 2010 3 41 00 PM Sample Data g Item 3 Target 1 000 g...

Страница 132: ...e 0 00 g Difference 0 0 Signature Verified by Company Name Address 1 Address 2 Address 3 Address 4 12 7 2010 4 00 00 PM Balance ID Balance Name EXPLORER User Name Project Name Application Peak Hold Sa...

Страница 133: ...Limit 0 00 Standard Deviation 0 02ml Number of Samples 2 2S 0 00 0 1S 0 00 0 1S Mean 1S 100 00 2 1S 0 00 0 2S 0 00 0 Sample Data ml 1 2 7 2 2 8 Signature Verified by Company Name Address 1 Address 2 A...

Страница 134: ...0 09 4 200 06 5 200 06 6 200 07 7 200 07 8 200 08 9 200 08 10 200 09 Signature Verified by Company Name Address 1 Address 2 Address 3 Address 4 12 7 2010 4 14 00 PM Balance ID Balance Name EXPLORER Us...

Страница 135: ...ID Signature Verified by End User Calibration Company Name Address 1 Address 2 Address 3 Address 4 12 6 2010 4 36 AM User Calibration Balance ID Balance Name EXPLORER User Name Project Name Calibratio...

Страница 136: ...3 2 1 1 In certain cases a Label field of up to 11 characters is included 2 Each field is followed by a single delimiting space ASCII 32 3 The Weight field is 9 right justified characters If the value...

Страница 137: ...on Unstable environment Perform calibration Move balance to suitable location Cannot calibrate Calibration Menu locked LFT set to on Unstable environment Incorrect calibration masses Turn Calibration...

Страница 138: ...18 10 stainless steel Weighing Pan 304 SST EX12001 EX24001 and EX35001 In use Cover Terminal plastic PET Draft Shield Glass plastic ABS Battery Specification Battery Type Lithium Ion 2600mAh 14 8V Ba...

Страница 139: ...stment Statistical Quality Control Stabilization time typical 8s Sensitivity Temperature Drift 0 80 ppm K Typical Operating Range USP u 0 10 k 2 0 02g 120g 0 02g 120g 0 02g 220g 0 02g 220g 0 02g 220g...

Страница 140: ...hin 2 seconds Within 3 seconds Sensitivity Temperature Drift PPM C 1 5 1 5 1 5 Typical Operating Range USP u 0 10 k 2 0 16g 120g 0 16g 220g 0 16g 320g Optimal Operating Range USP u 0 10 k 2 0 082g 120...

Страница 141: ...Within 1 5 seconds Sensitivity Temperature Drift PPM C 3 3 3 3 Typical Operating Range USP u 0 10 k 2 1 6g 220g 1 6g 420g 1 6g 620g 1 6g 1100g Optimal Operating Range USP u 0 10 k 2 0 82g 220g 0 82g...

Страница 142: ...time typical Within 1 second Sensitivity Temperature Drift PPM C 3 3 3 3 Typical Operating Range USP u 0 10 k 2 16g 2200g 16g 4200g 16g 6200g 16g 10200g Optimal Operating Range USP u 0 10 k 2 8 2g 220...

Страница 143: ...y Temperature Drift PPM C 5 3 Typical Operating Range USP u 0 10 k 2 160g 6200g 160g 10200g Optimal Operating Range USP u 0 10 k 2 82g 6200g 82g 10200g Display Full Color VGA Graphic LCD Display size...

Страница 144: ...hin 1 second Sensitivity Temperature Drift PPM C 5 Typical Operating Range USP u 0 10 k 2 160g 12000g 160g 24000g 160g 35001g Optimal Operating Range USP u 0 10 k 2 82g 12000g 82g 24000g 82g 35000g Di...

Страница 145: ...bled dimensions Figure 9 1 Explorer Semi Micro Balance Figure 9 2 Draft Shield models Figure 9 3 Non Draft Shield models 15 5 in 393 mm 3 9 in 98 mm 9 1 in 230 mm 9 1 in 230 mm 13 8 in 350 mm 15 5 in...

Страница 146: ...EN 144 EXPLORER BALANCES Figure 9 4 EX12001 EX24001 and EX35001 models 14 8 in 377 mm 4 7 in 120 mm 17 4 in 443 mm 12 1 in 1 9 in 48 mm 311 mm www com information itm com 1 800 561 8187...

Страница 147: ...ot Switch kit 83021086 Terminal Extension Cable kit 83021083 Density Determination kit 80253384 Tower Mount for Terminal 83021102 Draft Shield kit 83021084 Security Device 80850043 Printer SF 40A 3006...

Страница 148: ...ts Menu AutoCal GLP and GMP Data Menu Main Menu Communication Menu More Functions Library Level Assist User Management Zero Password Policy Tare Diagnostics Menu Units Factory Reset Menu Pre Tare Info...

Страница 149: ...ic Calibration Filter Level AutoCal Adjustment Auto Zero Tracking Span Calibration Auto Tare User Calibration Gross Indicator Calibration Test Legal For Trade Language Graduations Volume Date and Time...

Страница 150: ...N Weighing Formulation Parts Counting Differential Percent Weighing Density Determination Check Weighing Peak Hold Dynamic Weighing Ingredient Costing not available in semi micro models Filling Pipett...

Страница 151: ...Momme unit Gram unit Mesghal unit Kilogram unit Hong Kong Tael unit Carat unit Singapore Tael unit Ounce unit Taiwan Tael unit Ounce Troy unit Tical unit Pound unit Tola unit Pennyweight unit Baht un...

Страница 152: ...ame Balance Setup Reset RS 232 Standard Application Modes Reset USB Standard Weighing Units Reset Repeatability Test GLP and GMP Data Reset Number of Repeatability Tests Communication Reset Test Weigh...

Страница 153: ...25 characters the setting format is H x header line text and H x could get the corresponding header line in the balance Z Same as pressing Zero Key T Same as pressing Tare Key xT Establish a preset Ta...

Страница 154: ...de xC Set Checkweighing tolerance x Attention when x is a positive value it is used to set the tolerance value vice versa PC Print Checkweighing tolerance xCW Set Checkweighing weight tolerance x Atte...

Страница 155: ...Index Application Index Application 0 Weighing 11 Ingredient Adjustment 1 Parts Counting 12 Pipette Adjustment 2 Percent Weighing 13 SQC 3 Check Weighing 14 Fill Weight Variation 4 Dynamic Weighing 5...

Страница 156: ...ows XP or Windows 7 Windows 10 Available USB port Type A 4 pin female USB Connection The balance s USB port terminates with a mini USB A USB Cable type B male to type A male is required not supplied 1...

Страница 157: ...ance sends data every xx seconds PRINT Auto Print Off Only sends data when PRINT is pressed USB Baud Parity Handshake Set parameters to match printer or computer Operation USB OUTPUT With the Adapter...

Страница 158: ...racter when unstable or below min weight blank when stable space 1 character legend 10 character TOTAL hh mm ss time interval etc CR 1 character LF 1 character Output examples Note and _ indicate spac...

Страница 159: ...table CP Continuous print of weights SLP Auto print stable non zero weight only SLZP Auto print stable non zero weight and zero reading xP Auto print on 1 to 3600 second intervals x 1 to 3600 0P turns...

Страница 160: ...ility of the purchaser to ensure that all pertinent legal requirements are met Weighing Instruments verified at the place of manufacture bear the following supplementary metrology marking on the descr...

Страница 161: ...its for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commer...

Страница 162: ...dent or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than Ohaus In lieu of a pr...

Страница 163: ...www com information itm com 1 800 561 8187...

Страница 164: ...us Drive Suite 310 Parsippany NJ 07054 USA Tel 973 377 9000 Fax 973 944 7177 With offices worldwide www ohaus com 83021171 P N 83021171 M 2019 Ohaus Corporation all rights reserved www com information...

Отзывы: