EN-74
EXPLORER BALANCES
4.10.6 Measuring the Density of Porous Material Using Oil
To enable this feature, enter the Density Determination
Setup
menu, and set the following:
Density Type:
Solid
Liquid Type:
Water
Porous Material:
On
The
DENSITY DETERMINATION
– POROUS
Home screen
Main Display Line
Second Display Line
Reference Fields
Application Buttons
Functions Application Icon
Confirm the default values displayed (Water Temp) are correct.
To edit the default values, touch
Edit settings
.
The Settings screen appears.
Settings available:
Water Temp, Oil Density, Sample
Name, Sample ID, Batch ID, Clear All.
Note:
please refer to section 4.1.1 for details about setting
Sample Name, Sample ID, Batch ID and Clear All.
Functions Available:
Return to Application
The balance calculates water density based on the water
temperature value entered (look-up table).
Measure the actual water temperature using a precision
thermometer.
To adjust the Water Temperature or Oil Density values,
touch the
Water Temp
or
Oil Density
button.
Numeric input windows appear.
Key in the desired value,
then press
.
The display returns to the previous
screen with the new value highlighted.
To return to the Density Determination
home screen, touch
Return to Application
.
www.
.com