EN-32
Defender 6000 Indicators
IR Function
[
Ir.FUNC
]
Define the operation for the Infrared Radiation Sensor.
ZerO
= the scale performs zero operation equal to pressing the
Zero
button.
tare
= the scale performs tare operation equal to pressing the
button.
= the scale performs print operation equal to pressing the
button.
Display
= turns on the backlight.
Wave you hand before the
icon on the control panel to perform the operation you set.
Note:
This setting is only avaliable for i-DT61XWE model.
IR Adjust
[
Ir.AdJ
]
Define the response distance of the Infrared Radiation Sensor.
OFF
= the IR sensor is disabled.
LOW
= the sensor will respond within 50mm/2 inches.
HIGH
= the sensor will respond within 100mm/4 inches.
Note:
This setting is only avaliable for i-DT61XWE model.
Password Enable
[PwWd.En]
Define if you need password to enter menu. When is it turned on, you need to input the password each
time you enter menu.
OFF
= to disable password function.
ON
= to enable password function.
Password
[PwWd]
This sub-menu will only appear when Password Enable menu is selected to be on.
The display flashes with
_
, input the new password through the numeric keyboard. Press the
CLR
button
to change your inputs. Press the
Yes
button when you finish.
End Setup
[
End
]
Advance to the next menu or return to the top of the current menu.