Defender 6000 Indicators
EN-19
3.3
Counting Mode
Use this application to count samples of uniform weight.
3.3.1 Enter the Mode
To enter the mode:
1. Press and hold the
Mode
button until
COUNt
is displayed.
2. When the
Mode
button is released, the display shows
CLr.PwW
.
3. If you need to clear the stored APW of the last time, press the
Yes
button. Then go to
Establish an APW
section.
Note:
if the weight on the pan is larger than 1d, the display will show
CLr.PAN
until the weight is removed
from the pan.
4. If you need to recall the stored APW of the last time and continue to use it, press the
No
button to start
counting.
Note:
if no APW has been set before, step 3 and 4 will be omitted.
3.3.2 Establish an APW
To establish an APW:
1. Follow the previous step 4.
i-DT61PW:
The display shows the sample size
PUt.10
. To change it, short press the
No
button several times until you
see the value you want.
Note:
available sample size selections are 5, 10, 20, 50 and 100 (The default is 10).
i-DT61XWE:
The display flashes with the current sample size, such as 10 Pcs. To change it, input the new sample size
through the numeric keyboard. Do not press the
Yes
button until you finish the next step.
2. Place the specified quantity of samples on the pan and press the
Yes
button to capture the current stable
weight.
Note:
During the capture process, the display shows
- - - - -
. (i-DT61PW)
You can press the button to tare. The center of zero, PT or NET icons will light as appropriate.
If the APW is between 0.1d and 1d, the display shows
LO.rEF
for 1.5 seconds. Then it will start counting.
If the APW is less than 0.1d, the display shows
rEF.Err
for 1.5 seconds and then return to showing what
is displayed on step 1. Please replace the samples on the pan and press the
Yes
button to re-establish an
APW value.
3.3.3 Start Counting
1. Place parts on the pan and read the number. The number of pieces and the Pcs icon are displayed.
2. Press the
Function
button to temporarily display the APW.
APwW
is displayed for 0.5 seconds. Then the
APW value is displayed for 1.5 seconds using the current unit of measurement.
Note:
Please refer to
Check
in the
Weighing Mode
section for how to use Check in the Counting mode.
3.3.4 Application Settings
The application can be customized for user preferences. Please refer to
Application Settings
section in
Weighing Mode
for details about how to enter application settings.
The Counting Configurations are defined below (defaults in Bold).
Item
Available Settings
Comments
Count
(
COUNt
)