3. OPERATION 

3.1  Turning On/Off the Scale 

To turn the scale on, short press the 

ON/ZERO Off 

button. The scale performs a display test, momentarily 

displays the model/software version, and then enters the weighing mode. 

 

To turn the scale off, long press 

ON/ZERO Off

 button until OFF is displayed. 

3.2  Zero Operation 

By short pressing the 

ON/ZERO Off

 button, the scale display returns to zero. When adding additional mass, the 

ON/ZERO Off

 button may be used repeatedly until the full capacity of the scale is reached. When a container is 

used, the 

ON/ZERO Off

 button may be used to initiate Tare function. Additional mass may then be added as a 

NET weight. When removing both the sample and container from the scale, a negative value may be displayed. 

Zero the scale again before subsequent usage. 

NOTE

: The 

ON/ZERO

 

Off

 button will perform a true zero setting function when displayed values are 2% 

of full capacity. Above 2%, the range is limited to full capacity by subtraction. 

3.3  Changing the Units of Measure 

Short press the 

UNITS Cal

 button to display the next available measuring unit. 

3.4  Auto Shut-Off 

To extend battery life, the scale will automatically turn off after approximately four minutes of inactivity. This 

feature is only active during battery operation. 

3.5  Dynamic Weighing 

With the platform empty, short press the 

UNITS Cal

 button to cycle 

through the units. Dynamic mode is active when the symbol " • “is 

flashing. 

 

Place an item on the platform. The scale will show “-A-”. 

 

NOTE

: If using a container, press 

ON/ZERO Off

 to tare the weight and 

return to zero. 

 

With an item or a container on the platform, a short press of 

UNITS Cal

 

will start the averaging countdown from 5 seconds. 

 

The average weight is then displayed. The terminal • stops flashing 

when the process is complete. 

 

The display will hold until a button is pressed. 

 

To repeat the process, press 

UNITS Cal

 with an item or a container on 

the platform. 

 

 

To Exit, remove all weight from the platform or container and short 

press 

UNITS Cal 

button. 

 

3.6  Weight Alert

TM

 

With platform empty, press the 

UNITS Cal 

button until the lb unit displayed. 

 

The digits will start to flash if the weight on the platform is over 70 lb. 

If the weight on the pan is over 150 lb, the digits will stop flashing.  

 

NOTE

: Weigh Alert

TM

 is only available on i-C12P75 model. Weight Alert

TM

 is available only in lb mode without 

Dynamic Weighing. Make sure that the "lb" is turned On and the "•" dynamic weighing is turned Off. 

Содержание Courier 1000

Страница 1: ...Courier 1000 Instruction Manual ...

Страница 2: ......

Страница 3: ...REVISION HISTORY Date Version Description February 6 2023 A New manual ...

Страница 4: ......

Страница 5: ...pter Power 4 2 3 2 Battery Power 4 3 OPERATION 5 3 1 Turning On Off the Scale 5 3 2 Zero Operation 5 3 3 Changing the Units of Measure 5 3 4 Auto Shut Off 5 3 5 Dynamic Weighing 5 3 6 Weight AlertTM 5 4 CALIBRATION 6 4 1 Span Calibration 6 4 2 Calibration Locking 7 5 MAINTENANCE 8 5 1 Cleaning 8 5 2 Troubleshooting 8 5 3 Service Information 8 6 TECHNICAL DATA 9 6 1 Specifications 9 6 2 Drawings an...

Страница 6: ...hat it is difficult to reach the power connection Make sure that the power cord does not pose a potential obstacle or tripping hazard Operate the equipment only under ambient conditions specified in these instructions The equipment is for indoor use only Do not operate the equipment in wet hazardous or unstable environments Do not allow liquids to enter the equipment Do not load the equipment abov...

Страница 7: ...t side of the display indicates a low battery condition When first displayed approximately 12 hours of operation remain When the battery is fully depleted the scale will momentarily display Lo bAt and shut off 2 Large 7 Segment Numeric Characters Weight mass values are shown using 4 digits including negative sign and decimal places Prompts for calibration and possible error conditions are also sho...

Страница 8: ...available Connect the AC adapter plug to the input jack Then plug the AC adapter into a properly grounded power outlet Figure 2 1 Power Connection 2 3 2 Battery Power Open the battery cover on the bottom of the scale and install 3 x C LR14 batteries into the compartment Orient the batteries as shown on the inside of the compartment Close the battery cover NOTE Batteries are not included in the pac...

Страница 9: ...end battery life the scale will automatically turn off after approximately four minutes of inactivity This feature is only active during battery operation 3 5 Dynamic Weighing With the platform empty short press the UNITS Cal button to cycle through the units Dynamic mode is active when the symbol is flashing Place an item on the platform The scale will show A NOTE If using a container press ON ZE...

Страница 10: ...t Model Calibration weight kg lb i C12P9 9 20 i C12P20 20 40 i C12P75 50 100 4 1 Span Calibration Span Calibration uses two points to adjust the scale The first point is the zero value where there is no weight on the scale The second point is the Span value where a calibration mass is placed on the scale To select the weighing unit to be used for calibration press the UNITS Cal button until the co...

Страница 11: ...screws at the bottom Slide the switch on the PCB marked SW3 to the position marked CAL LOCK Reassemble the cover If required place tamper evident sealing labels over the screw holes or over the housing edges If later recalibration is required the cover will need to be reopened and the lock switch returned to the original position NOTE When the lock switch is set to the CAL LOCK position the power ...

Страница 12: ...teries 4 Service required Poor accuracy 1 Improper calibration 2 Unstable environment 1 Perform calibration 2 Move scale to suitable location Unable to calibrate 1 CAL LOCK is set to ON 2 Unstable environment 3 Incorrect calibration mass 1 Set CAL LOCK to OFF 2 Move the scale to suitable location 3 Use correct calibration mass Scale displays Lo bAt Low battery Connect the scale to power outlet Rep...

Страница 13: ... 02 lb 75 kg x 0 05 kg 165 lb x 0 1 lb Weighing Units kg lb Modes Weighing Dynamic Weighing Weight AlertTM Weigh AlertTM is only available when lb is activated on i C12P75 model Keypad 2 button mechanical keys Calibration Weights 9 kg 20 lb 20 kg 40 lb 50 kg 100 lb Display 4 digit 7 segment 20mm 0 8 in characters Tare range To capacity by subtraction Stabilization Time 3 seconds Maximum Overload 1...

Страница 14: ...rical and Electronic Equipment Regulations 2012 UK Electromagnetic Compatibility Regulations 2016 and Electrical Equipment Safety Regulations 2016 The UK Declaration of Conformity is available online at www ohaus com uk declarations This product complies with the EU Directive 2012 19 EU WEEE Please dispose of this product in accordance with local regulations at the collecting point specified for e...

Страница 15: ...e harmful interference in which case the user will be required to correct the interference at his own expense Changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment LIMITED WARRANTY OHAUS products are warranted against defects in materials and workmanship from the date of delivery through the duration of the ...

Страница 16: ......

Страница 17: ...offices worldwide Con oficinas en todo el mundo Avec des bureaux partoutdans le monde MitBürosweltweit Con uffici in tuttoilmondo www ohaus com 30746010 P N 30746010 A 2023 Ohaus Corporation all rights reserved todos los derechos reservados tous droits réservés alle Rechte vorbehalten tutti i diritti riservati ...

Отзывы: