OHAUS Courier 1000 Скачать руководство пользователя страница 12

EN-8 

 

Courier 1000 

5. MAINTENANCE 

Caution

: Before cleaning, turn off the scale, and remove the AC adapter. 

5.1  Cleaning 

 

WARNING

: Electric Shock Hazard. Disconnect the equipment from the power supply before cleaning.  

Make sure that no liquid enters the interior of the instrument. 

 

Attention:

 Do not use solvents, harsh chemicals, ammonia or abrasive cleaning agents. 

 

The housing may be cleaned with a cloth dampened with a mild detergent if necessary. 

5.2  Troubleshooting 

The following table lists common problems and possible causes and remedies.  If the problem persists, contact 

Ohaus or your authorized dealer. 

 

Table 5-1 Troubleshooting 

Symptom 

Possible Cause(s) 

Remedy 

Unable to turn on the scale 

1.  Power cord not plugged in or 

properly connected. 

2.  Power outlet not supplying power. 

3.  Batteries out of power. 

4.  Other failure. 

1.  Check power cord connections. Make 

sure power cord is plugged in properly 

into the power outlet. 

2.  Check the power source. 

3.  Change batteries. 

4.  Service required. 

Poor accuracy 

1.  Improper calibration.  

2.  Unstable environment. 

1.  Perform calibration.  

2.  Move scale to suitable location. 

Unable to calibrate 

1.  CAL LOCK is set to ON. 

2.  Unstable environment. 

3.  Incorrect calibration mass. 

1.  Set CAL LOCK to OFF. 

2.  Move the scale to suitable location. 

3.  Use correct calibration mass. 

Scale displays “Lo bAt” 

Low battery 

Connect the scale to power outlet. 

Replace batteries. 

Scale displays “Lo Line”  Low Line voltage 

Verify source voltage. 

Scale displays “E” 

Weight on platform exceeds capacity  Remove load on scale. 

Scale displays “CALE” 

Calibration failure

 

Use the correct calibration weight. 

Scale displays “UnSt” 

Scale is unstable at power up 

Relocate scale to a more stable environment. 

 

5.3  Service Information 

If the troubleshooting section does not resolve or describe your problem, contact your authorized Ohaus 

service agent. For service assistance or technical support in the United States call toll-free 1-800-526-0659 

between 8.00 AM and 5:00 PM EST. An Ohaus product service specialist will be available to help. Outside the 

USA, please visit our website, www.ohaus.com to locate the Ohaus office nearest you. 

 

 

Содержание Courier 1000

Страница 1: ...Courier 1000 Instruction Manual ...

Страница 2: ......

Страница 3: ...REVISION HISTORY Date Version Description February 6 2023 A New manual ...

Страница 4: ......

Страница 5: ...pter Power 4 2 3 2 Battery Power 4 3 OPERATION 5 3 1 Turning On Off the Scale 5 3 2 Zero Operation 5 3 3 Changing the Units of Measure 5 3 4 Auto Shut Off 5 3 5 Dynamic Weighing 5 3 6 Weight AlertTM 5 4 CALIBRATION 6 4 1 Span Calibration 6 4 2 Calibration Locking 7 5 MAINTENANCE 8 5 1 Cleaning 8 5 2 Troubleshooting 8 5 3 Service Information 8 6 TECHNICAL DATA 9 6 1 Specifications 9 6 2 Drawings an...

Страница 6: ...hat it is difficult to reach the power connection Make sure that the power cord does not pose a potential obstacle or tripping hazard Operate the equipment only under ambient conditions specified in these instructions The equipment is for indoor use only Do not operate the equipment in wet hazardous or unstable environments Do not allow liquids to enter the equipment Do not load the equipment abov...

Страница 7: ...t side of the display indicates a low battery condition When first displayed approximately 12 hours of operation remain When the battery is fully depleted the scale will momentarily display Lo bAt and shut off 2 Large 7 Segment Numeric Characters Weight mass values are shown using 4 digits including negative sign and decimal places Prompts for calibration and possible error conditions are also sho...

Страница 8: ...available Connect the AC adapter plug to the input jack Then plug the AC adapter into a properly grounded power outlet Figure 2 1 Power Connection 2 3 2 Battery Power Open the battery cover on the bottom of the scale and install 3 x C LR14 batteries into the compartment Orient the batteries as shown on the inside of the compartment Close the battery cover NOTE Batteries are not included in the pac...

Страница 9: ...end battery life the scale will automatically turn off after approximately four minutes of inactivity This feature is only active during battery operation 3 5 Dynamic Weighing With the platform empty short press the UNITS Cal button to cycle through the units Dynamic mode is active when the symbol is flashing Place an item on the platform The scale will show A NOTE If using a container press ON ZE...

Страница 10: ...t Model Calibration weight kg lb i C12P9 9 20 i C12P20 20 40 i C12P75 50 100 4 1 Span Calibration Span Calibration uses two points to adjust the scale The first point is the zero value where there is no weight on the scale The second point is the Span value where a calibration mass is placed on the scale To select the weighing unit to be used for calibration press the UNITS Cal button until the co...

Страница 11: ...screws at the bottom Slide the switch on the PCB marked SW3 to the position marked CAL LOCK Reassemble the cover If required place tamper evident sealing labels over the screw holes or over the housing edges If later recalibration is required the cover will need to be reopened and the lock switch returned to the original position NOTE When the lock switch is set to the CAL LOCK position the power ...

Страница 12: ...teries 4 Service required Poor accuracy 1 Improper calibration 2 Unstable environment 1 Perform calibration 2 Move scale to suitable location Unable to calibrate 1 CAL LOCK is set to ON 2 Unstable environment 3 Incorrect calibration mass 1 Set CAL LOCK to OFF 2 Move the scale to suitable location 3 Use correct calibration mass Scale displays Lo bAt Low battery Connect the scale to power outlet Rep...

Страница 13: ... 02 lb 75 kg x 0 05 kg 165 lb x 0 1 lb Weighing Units kg lb Modes Weighing Dynamic Weighing Weight AlertTM Weigh AlertTM is only available when lb is activated on i C12P75 model Keypad 2 button mechanical keys Calibration Weights 9 kg 20 lb 20 kg 40 lb 50 kg 100 lb Display 4 digit 7 segment 20mm 0 8 in characters Tare range To capacity by subtraction Stabilization Time 3 seconds Maximum Overload 1...

Страница 14: ...rical and Electronic Equipment Regulations 2012 UK Electromagnetic Compatibility Regulations 2016 and Electrical Equipment Safety Regulations 2016 The UK Declaration of Conformity is available online at www ohaus com uk declarations This product complies with the EU Directive 2012 19 EU WEEE Please dispose of this product in accordance with local regulations at the collecting point specified for e...

Страница 15: ...e harmful interference in which case the user will be required to correct the interference at his own expense Changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment LIMITED WARRANTY OHAUS products are warranted against defects in materials and workmanship from the date of delivery through the duration of the ...

Страница 16: ......

Страница 17: ...offices worldwide Con oficinas en todo el mundo Avec des bureaux partoutdans le monde MitBürosweltweit Con uffici in tuttoilmondo www ohaus com 30746010 P N 30746010 A 2023 Ohaus Corporation all rights reserved todos los derechos reservados tous droits réservés alle Rechte vorbehalten tutti i diritti riservati ...

Отзывы: