1. INTRODUCTION 

This manual contains installation, operation, and maintenance instructions for Courier 1000. This instruction 

manual helps you to install and use this scale easily. Therefore, you must read it carefully before installation 

and operation.  

1.1.  Safety Precautions 

Definition of Signal Warnings and Symbols 

Safety notes are marked with signal words and warning symbols. These show safety issues and warnings. 

Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results. 

Signal Words 

WARNING 

For a hazardous situation with medium risk, possibly resulting in severe injuries or 

death if not avoided. 

CAUTION 

For a hazardous situation with low risk, resulting in damage to the device or the 

property or in loss of data, or minor or medium injuries if not avoided. 

ATTENTION 

For important information about the product. May lead to equipment damage if not 

avoided. 

NOTE 

For useful information about the product. 

Warning Symbols 

 

General hazard 

 

Explosion hazard 
 

 

Electrical shock hazard 

Safety Precautions 

CAUTION: 

Read all safety warnings before installing, making connections, or servicing this 

equipment. Failure to comply with these warnings could result in personal injury and/or property 

damage. Retain all instructions for future reference. 

  Before connecting power, verify that the AC adapter’s input voltage range and plug type are 

compatible with the local AC mains power supply. 

  Do not position the equipment such that it is difficult to reach the power connection. 

  Make sure that the power cord does not pose a potential obstacle or tripping hazard. 

  Operate the equipment only under ambient conditions specified in these instructions. 

  The equipment is for indoor use only. 

  Do not operate the equipment in wet, hazardous, or unstable environments. 

  Do not allow liquids to enter the equipment. 

  Do not load the equipment above its rated capacity. 

  Do not drop loads on the platform. 

  Do not place the equipment upside down on the platform. 

  Use only approved accessories and peripherals. 

  Disconnect the equipment from the power supply when cleaning. 

  Service should only be performed by authorized personnel. 

 

WARNING: Never work in an environment subject to explosion hazards! The housing of the 

instrument is not gas tight. (Explosion hazard due to spark formation, corrosion caused by the 

ingress of gases). 

 

WARNING: Electrical shock hazards exist within the housing. The housing should only be opened by 

authorized and qualified personnel. Remove all power connections to the unit before opening. 

1.2.  Intended Use 

This instrument is intended for use in pharmacies, schools, businesses and light industry. It must only be used 

for measuring the parameters described in these operating instructions. Any other type of use and operation 

beyond the limits of technical specifications, without written consent from OHAUS, is considered as not 

intended. This instrument complies with current industry standards and the recognized safety regulations; 

however, it can constitute a hazard in use. If the instrument is not used according to these operating 

instructions, the intended protection provided by the instrument may be impaired. 

 

Содержание Courier 1000

Страница 1: ...Courier 1000 Instruction Manual ...

Страница 2: ......

Страница 3: ...REVISION HISTORY Date Version Description February 6 2023 A New manual ...

Страница 4: ......

Страница 5: ...pter Power 4 2 3 2 Battery Power 4 3 OPERATION 5 3 1 Turning On Off the Scale 5 3 2 Zero Operation 5 3 3 Changing the Units of Measure 5 3 4 Auto Shut Off 5 3 5 Dynamic Weighing 5 3 6 Weight AlertTM 5 4 CALIBRATION 6 4 1 Span Calibration 6 4 2 Calibration Locking 7 5 MAINTENANCE 8 5 1 Cleaning 8 5 2 Troubleshooting 8 5 3 Service Information 8 6 TECHNICAL DATA 9 6 1 Specifications 9 6 2 Drawings an...

Страница 6: ...hat it is difficult to reach the power connection Make sure that the power cord does not pose a potential obstacle or tripping hazard Operate the equipment only under ambient conditions specified in these instructions The equipment is for indoor use only Do not operate the equipment in wet hazardous or unstable environments Do not allow liquids to enter the equipment Do not load the equipment abov...

Страница 7: ...t side of the display indicates a low battery condition When first displayed approximately 12 hours of operation remain When the battery is fully depleted the scale will momentarily display Lo bAt and shut off 2 Large 7 Segment Numeric Characters Weight mass values are shown using 4 digits including negative sign and decimal places Prompts for calibration and possible error conditions are also sho...

Страница 8: ...available Connect the AC adapter plug to the input jack Then plug the AC adapter into a properly grounded power outlet Figure 2 1 Power Connection 2 3 2 Battery Power Open the battery cover on the bottom of the scale and install 3 x C LR14 batteries into the compartment Orient the batteries as shown on the inside of the compartment Close the battery cover NOTE Batteries are not included in the pac...

Страница 9: ...end battery life the scale will automatically turn off after approximately four minutes of inactivity This feature is only active during battery operation 3 5 Dynamic Weighing With the platform empty short press the UNITS Cal button to cycle through the units Dynamic mode is active when the symbol is flashing Place an item on the platform The scale will show A NOTE If using a container press ON ZE...

Страница 10: ...t Model Calibration weight kg lb i C12P9 9 20 i C12P20 20 40 i C12P75 50 100 4 1 Span Calibration Span Calibration uses two points to adjust the scale The first point is the zero value where there is no weight on the scale The second point is the Span value where a calibration mass is placed on the scale To select the weighing unit to be used for calibration press the UNITS Cal button until the co...

Страница 11: ...screws at the bottom Slide the switch on the PCB marked SW3 to the position marked CAL LOCK Reassemble the cover If required place tamper evident sealing labels over the screw holes or over the housing edges If later recalibration is required the cover will need to be reopened and the lock switch returned to the original position NOTE When the lock switch is set to the CAL LOCK position the power ...

Страница 12: ...teries 4 Service required Poor accuracy 1 Improper calibration 2 Unstable environment 1 Perform calibration 2 Move scale to suitable location Unable to calibrate 1 CAL LOCK is set to ON 2 Unstable environment 3 Incorrect calibration mass 1 Set CAL LOCK to OFF 2 Move the scale to suitable location 3 Use correct calibration mass Scale displays Lo bAt Low battery Connect the scale to power outlet Rep...

Страница 13: ... 02 lb 75 kg x 0 05 kg 165 lb x 0 1 lb Weighing Units kg lb Modes Weighing Dynamic Weighing Weight AlertTM Weigh AlertTM is only available when lb is activated on i C12P75 model Keypad 2 button mechanical keys Calibration Weights 9 kg 20 lb 20 kg 40 lb 50 kg 100 lb Display 4 digit 7 segment 20mm 0 8 in characters Tare range To capacity by subtraction Stabilization Time 3 seconds Maximum Overload 1...

Страница 14: ...rical and Electronic Equipment Regulations 2012 UK Electromagnetic Compatibility Regulations 2016 and Electrical Equipment Safety Regulations 2016 The UK Declaration of Conformity is available online at www ohaus com uk declarations This product complies with the EU Directive 2012 19 EU WEEE Please dispose of this product in accordance with local regulations at the collecting point specified for e...

Страница 15: ...e harmful interference in which case the user will be required to correct the interference at his own expense Changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment LIMITED WARRANTY OHAUS products are warranted against defects in materials and workmanship from the date of delivery through the duration of the ...

Страница 16: ......

Страница 17: ...offices worldwide Con oficinas en todo el mundo Avec des bureaux partoutdans le monde MitBürosweltweit Con uffici in tuttoilmondo www ohaus com 30746010 P N 30746010 A 2023 Ohaus Corporation all rights reserved todos los derechos reservados tous droits réservés alle Rechte vorbehalten tutti i diritti riservati ...